Basics of LC/MS

Basics of LC/MS

Basics of LC/MS Primer Contents Why Liquid Chromatography/Mass Spectrometry? 4 Instrumentation 6 Ion Sources 6 Electrospray ionization 7 Atmospheric pressure chemical ionization 8 Atmospheric pressure photoionization 9 Mass Analyzers 10 Quadrupole 10 Time-of-flight 11 Ion trap 11 Fourier transform-ion cyclotron resonance (FT-ICR) 12 Collision-Induced Dissociation and Multiple-Stage MS 13 CID in single-stage MS 14 CID and multiple-stage MS 14 Adapting LC Methods 16 Sample preparation 17 Ionization chemistry 17 Capillary LC/MS and CE/MS 19 Applications 20 Molecular Weight Determination 20 Differentiation of similar octapeptides 20 Determining the molecular weight of green fluorescent protein 21 Structural Determination 22 Structural determination of ginsenosides using MSn analysis 22 Pharmaceutical Applications 24 Rapid chromatography of benzodiazepines 24 Detection of degradation products for salbutamol 24 Identification of bile acid metabolites 25 Biochemical Applications 26 Rapid protein identification using capillary LC/MS/MS and database searching 26 Clinical Applications 28 High-sensitivity detection of trimipramine and thioridazine 28 Food Applications 28 Identification of aflatoxins in food 28 3 Determination of vitamin D3 in poultry feed supplements using MS 30 Environmental Applications 32 Detection of phenylurea herbicides 32 Detection of low levels of carbaryl in food 32 CE/MS Applications 34 Analysis of peptides using CE/MS/MS 34 Why Liquid Chromatography/ Mass spectrometers also generate three- Mass Spectrometry? dimensional data. In addition to signal strength, they generate mass spectral Liquid chromatography is a fundamental data that can provide valuable information separation technique in the life sciences about the molecular weight, structure, and related fields of chemistry. Unlike gas identity, quantity, and purity of a sample. chromatography, which is unsuitable for Mass spectral data add specificity that nonvolatile and thermally fragile molecules, increases confidence in the results of both liquid chromatography can safely qualitative and quantitative analyses. separate a very wide range of organic compounds, from small-molecule drug R metabolites to peptides and proteins. Traditional detectors for CH CH CH NHCOCH liquid chromatography 2 2 2 3 CH2CH2CH2OH include refractive index, CH2CH2CH(NH2)COOH electrochemical, fluores- cence, and ultraviolet-visible (UV-Vis) detectors. Some of these generate two- dimensional data; that is, data representing signal strength as a function of time. Others, including fluorescence and diode- array UV-Vis detectors, generate three-dimensional data. Three-dimensional data include not only signal strength but spectral data for each point in time. % Rel. Abundance% Rel. Abundance % Rel. m/z Figure 1. Two-dimensional abundance data and three-dimensional mass spectral data from a mass spectrometer 4 For most compounds, a mass spectrometer is Some mass spectrometers have the ability to more sensitive and far more specific than all perform multiple steps of mass spectrometry other LC detectors. It can analyze compounds on a single sample. They can generate a that lack a suitable chromophore. It can also mass spectrum, select a specific ion from identify components in unresolved chromato- that spectrum, fragment the ion, and generate graphic peaks, reducing the need for perfect another mass spectrum; repeating the entire chromatography. cycle many times. Such mass spectrometers can literally deconstruct a complex molecule Mass spectral data complements data from piece by piece until its structure is determined. other LC detectors. While two compounds may have similar UV spectra or similar mass spectra, it is uncommon for them to have both. The two orthogonal sets of data can be used to confidently identify, confirm, and quantify compounds. 1200000 800000 MS TIC 400000 4000 m/z 2000 648 12000 8000 m/z 4000 376 300 200 m/z 100 1325 5 10152025303540min. Figure 2. Identification of three components in a chromatographically unresolved peak 5 Instrumentation molecules from the analyte molecules (direct liquid inlet, thermospray) or did so before ion- Mass spectrometers work by ionizing mole- ization (particle beam). The analyte molecules cules and then sorting and identifying the ions were then ionized in the mass spectrometer according to their mass-to-charge (m/z) ratios. under vacuum, often by traditional electron Two key components in this process are the ionization. These approaches were successful ion source, which generates the ions, and the only for a very limited number of compounds. mass analyzer, which sorts the ions. Several different types of ion sources are commonly The introduction of atmospheric pressure used for LC/MS. Each is suitable for different ionization (API) techniques greatly expanded classes of compounds. Several different types the number of compounds that can be suc- of mass analyzers are also used. Each has cessfully analyzed by LC/MS. In atmospheric advantages and disadvantages depending on pressure ionization, the analyte molecules the type of information needed. are ionized first, at atmospheric pressure. The analyte ions are then mechanically Ion Sources and electrostatically separated from neutral molecules. Common atmospheric pressure Much of the advancement in LC/MS over the ionization techniques are: last ten years has been in the development of ion sources and techniques that ionize the • Electrospray ionization (ESI) analyte molecules and separate the resulting • Atmospheric pressure chemical ionization ions from the mobile phase. (APCI) • Atmospheric pressure photoionization (APPI) Earlier LC/MS systems used interfaces that either did not separate the mobile phase Figure 3. Applications of various 100,000 LC/MS ionization techniques Electrospray Ionization 10,000 1000 APPI APCI Molecular Weight 100 10 Nonpolar Very Polar Polarity 6 Electrospray ionization Some gas-phase reactions, mostly proton transfer and charge exchange, can also Electrospray relies in part on chemistry to gen- occur between the time ions are ejected erate analyte ions in solution before the analyte from the droplets and the time they reach reaches the mass spectrometer. The LC eluent the mass analyzer. is sprayed (nebulized) into a chamber at atmos- pheric pressure in the presence of a strong Electrospray is especially useful for analyzing electrostatic field and heated drying gas. large biomolecules such as proteins, peptides, and oligonucleotides, Ions but can also analyze Nebulizer gas smaller molecules like benzodiaze- pines and sulfated Heated nitrogen drying gas conjugates. Solvent spray Large molecules + + + often acquire more + + + + + + + + + + + + + + than one charge. Thanks to this multiple charging, Dielectric capillary entrance electrospray can be used to analyze molecules as large Figure 4. Electrospray ion source as 150,000 u even though the mass range (or more accurately The electrostatic field causes further mass-to-charge range) for a typical LC/MS dissociation of the analyte molecules. instruments is around 3000 m/z. For example: The heated drying gas causes the solvent 100,000 u / 10 z = 1,000 m/z in the droplets to evaporate. As the droplets shrink, the charge concentration in the When a large molecule acquires many charges, droplets increases. Eventually, the repulsive a mathematical process called deconvolution force between ions with like charges exceeds is often used to determine the actual molecular the cohesive forces and ions are ejected weight of the analyte. (desorbed) into the gas phase. These ions are attracted to and pass through a capillary sampling orifice into the mass analyzer. Evaporation Analyte ion ejected Figure 5. Desorption of ions from solution + + + + - +++ + + + ++- ++ - - + - -+ + - -- + + - -+ +- -++ - - - + +--+ ++ + ++ + ++ + -- + + ++-+ 7 Atmospheric pressure APCI is applicable to a wide range of polar chemical ionization and nonpolar molecules. It rarely results in multiple charging so it is typically used for In APCI, the LC eluent is sprayed through molecules less than 1,500 u. Due to this, a heated (typically 250°C – 400°C) vaporizer and because it involves high temperatures, at atmospheric pressure. The heat vaporizes APCI is less well-suited than electrospray the liquid. The resulting gas-phase solvent for analysis of large biomolecules that may molecules are ionized by electrons discharged be thermally unstable. APCI is used with from a corona needle. The solvent ions then normal-phase chromatography more often transfer charge to the analyte molecules than electrospray is because the analytes are through chemical reactions (chemical ioniza- usually nonpolar. tion). The analyte ions pass through a capillary sampling orifice into the mass analyzer. Figure 6. APCI ion source HPLC inlet Nebulizer (sprayer) Nebulizing gas Vaporizer (heater) Drying gas + + + + + + + + + + + + + + + Corona discharge needle Capillary 8 Atmospheric pressure photoionization APPI is applicable to many of the same compounds that are typically analyzed by Atmospheric pressure photoionization (APPI) APCI. It shows particular promise in two for LC/MS is a relatively new technique. As applications, highly nonpolar compounds in APCI, a vaporizer converts the LC eluent to and low flow rates (<100 µl/min), where APCI the gas phase. A discharge lamp generates sensitivity is sometimes reduced. photons in a narrow range of ionization energies. The range of energies is carefully In all cases, the nature of the analyte(s) chosen to ionize as many analyte

View Full Text

Details

  • File Type
    pdf
  • Upload Time
    -
  • Content Languages
    English
  • Upload User
    Anonymous/Not logged-in
  • File Pages
    36 Page
  • File Size
    -

Download

Channel Download Status
Express Download Enable

Copyright

We respect the copyrights and intellectual property rights of all users. All uploaded documents are either original works of the uploader or authorized works of the rightful owners.

  • Not to be reproduced or distributed without explicit permission.
  • Not used for commercial purposes outside of approved use cases.
  • Not used to infringe on the rights of the original creators.
  • If you believe any content infringes your copyright, please contact us immediately.

Support

For help with questions, suggestions, or problems, please contact us