EANM Guidelins on Routine Quality Control Recommendations for Nuclear

EANM Guidelins on Routine Quality Control Recommendations for Nuclear

Eur J Nucl Med Mol Imaging (2010) 37:662–671 DOI 10.1007/s00259-009-1347-y GUIDELINES Routine quality control recommendations for nuclear medicine instrumentation On behalf of the EANM Physics Committee: Ellinor Busemann Sokole & Anna Płachcínska & Alan Britten With contribution from the EANM Working Group on Nuclear Medicine Instrumentation Quality Control: Maria Lyra Georgosopoulou & Wendy Tindale & Rigobert Klett Published online: 4 February 2010 # EANM 2010 Keywords Quality control . Quality assurance . Nuclear Introduction medicine instrumentation . Gamma camera . SPECT. PET. CT. Radionuclide calibrator . Thyroid uptake probe . These recommendations cover routine quality control (QC) Nonimaging intraoperative probe . Gamma counting system . of instrumentation used within a nuclear medicine depart- Radiation monitors . Preclinical PET ment. Routine QC testing starts after installation of the instrument, and after acceptance testing, and continues on a regular basis throughout its lifetime. Additional periodic tests may be carried out to provide more in-depth testing. Recommendations for acceptance testing are covered in a separate document. These recommendations must be con- sidered in the light of any national guidelines and legislation, E. Busemann Sokole (*) which must be followed. The recommendations cover the Department of Nuclear Medicine, Academic Medical Centre, types of tests to be performed, and suggested frequencies, Amsterdam, The Netherlands e-mail: [email protected] but they do not specify the protocols to be followed, which are available from other reference sources quoted. A. Płachcínska Department of Nuclear Medicine, Medical University, Łódz, Poland Acceptance and reference tests A. Britten Department of Medical Physics and Bioengineering, St. George’s Healthcare NHS Trust, After installation, and before it is put into clinical use, a London, UK nuclear medicine instrument must undergo thorough and careful acceptance testing, the aim being to verify that the M. Lyra Georgosopoulou instrument performs according to its specifications and its Department of Radiology, Radiation Physics Unit, University of Athens, clinical purpose. Each instrument is supplied with a set of Athens, Greece basic specifications. These have been produced by the manufacturer according to standard test procedures, which W. Tindale should be traceable to standard protocols, such as the Medical Imaging and Medical Physics, – Sheffield Teaching Hospitals, NEMA and IEC performance standards [1 4, 11, 17, 30, Sheffield, UK 37]. By following such standard protocols in the clinical setting, with support from the vendor for supplying R. Klett phantoms and software where necessary, specifications Nuclear Medicine Practice, Faculty of Medicine, University of Giessen, can be verified and baseline performance data created. Giessen, Germany Additional tests are usually also needed in order to more Eur J Nucl Med Mol Imaging (2010) 37:662–671 663 thoroughly test individual components of an instrument. nuclear medicine department are presented in Tables 1, 2, 3, These acceptance test results form the reference data for 4, 5, 6, 7, 8 and 9. The tables give briefly the purpose of the future QC tests, and some may be repeated periodically, test, suggested frequency, and a comment. These are not such as at half-yearly or yearly intervals, or whenever a intended to supersede national guidelines or national major service or component change has been carried out. regulations, which should always be followed. The test frequencies given should be followed and adjusted accord- ing to observations of instrument stability and environmen- Routine tests tal stability (power supply, temperature and humidity). Where the frequency is indicated as a dual time period, e.g. Once the instrument has been accepted for clinical use, its weekly/monthly, it is recommended that tests are carried performance needs to be tested routinely with simple QC out at the shorter time period, and that the test frequency is procedures that are sensitive to changes in performance. Tests only reduced if regular testing gives evidence that the must be performed by appropriately qualified and trained system is stable. There is also a note in the legend to each staff, and detailed local operating procedures should be table giving the general type of technology that the written for this routine work. All test results must be recorded recommendations apply to. Specific systems may employ and monitored for variations, and appropriate actions taken technologies that need other checks, as specified by the when changes are observed. The QC tests are an important manufacturer. part of the routine work, and sufficient equipment time and It is a principle of radiation safety and patient protection staff time must be allocated for routine QC. to ensure that before the administration of a radiopharma- ceutical there is suitable equipment to complete the procedure, and that the equipment is functioning correctly. Action thresholds, follow-up, record keeping The correct basic function of the device must always be and review, and monitoring checked after any engineering intervention for repair or calibration, and appropriate engineer’s checks may satisfy Records of test results should be kept in a log-book or this need. digital record. Immediate review of the QC results is Note: The clocks within the department, within all essential, comparing them to the usual values obtained, and instruments and all computers must be synchronized and to the action thresholds for that test. Action thresholds checked daily/at least weekly. This is an essential require- should be set locally, taking into account the manufacturer’s ment for accurate activity administration and quantitative recommendations, and other professional guidance. Action data analysis. thresholds should be set to maintain the system within the specification value of a parameter. When action thresholds are reached a decision needs to be made about whether the Procedures and references instrument is fit for use. The QC operating procedure should make clear the actions to be taken when an action Test procedures are not included in this document, but can threshold is reached or exceeded, and who is responsible be found by reference to standard descriptions (e.g. [1–4]) for the decision to use the equipment. and national and international protocols. Any manufacturer- The follow-up actions taken when test results are supplied test procedures, test phantoms and software should unsatisfactory and solutions to a problem must be recorded. also be taken into account when writing local operating Such a record may assist troubleshooting if a similar procedures. Some essential literature references are given. problem occurs. They should be consulted for detailed generic test proce- dures and advice regarding test evaluation, action thresh- olds and follow-up. It is recommended that each department Routine test recommendations and test frequency create detailed written routine QC test procedures specific for each instrument in use. By following standard QC Recommendations for basic routine QC tests that should be procedures, consistency in test results can be assured and scheduled and carried out for the main instruments of the trends monitored. 664 Eur J Nucl Med Mol Imaging (2010) 37:662–671 Table 1 Routine QC tests for a gamma camera: planar, whole-body, SPECT and SPECT/CT. Equipment type: scintillation Anger gamma camera Test Purpose Frequency Comments GC1. Physical inspection To check collimator and detector Daily Inspect for mechanical and other defects that may head mountings, and to check for compromise safety of patient or staff; if any damage to the collimator collimator damage is detected or suspected, immediately perform a high-count extrinsic uniformity test GC2. Collimator touch pad To test that the touch pads and Daily Both the collimator touch pads and gantry and gantry emergency stop emergency stops are functioning emergency stop must function if there is an unexpected collision with the patient or an obstacle during motion; the touch pads must be checked each time the collimators are changed GC3. Energy window setting To check and centre the preset Daily The test is intended to check the correct 99mTc for 99mTc energy window on the 99mTc energy window photopeak GC4. Energy window setting – To test that preset energy windows Daily when used Frequency of the test should be adapted to the other radionuclides to be are properly centred for the particular camera and frequency of use of the used energies of other clinically used radionuclides radionuclides GC5. Background count rate To detect radioactive Daily The background count rate should be stable under contamination/excess electronic constant measuring conditions noise GC6. Intrinsic/extrinsic To test the response to a spatially Daily Visually inspect either an intrinsic or extrinsic uniformity and sensitivity uniform flux of 99mTc (or 57Co) (whichever is most convenient) low count for 99mTc (or 57Co) – visual photons, for uniformity and uniformity acquisition; if intrinsic method is overall sensitivity selected, each collimator must be checked periodically by an extrinsic uniformity test (preferably with high-count acquisition – see next test); record the cps/MBq to check and monitor sensitivity GC7. Intrinsic/extrinsic To monitor the trend in uniformity Weekly/monthly The most convenient method should be selected; uniformity and sensitivity with quantitative uniformity monitor uniformity

View Full Text

Details

  • File Type
    pdf
  • Upload Time
    -
  • Content Languages
    English
  • Upload User
    Anonymous/Not logged-in
  • File Pages
    10 Page
  • File Size
    -

Download

Channel Download Status
Express Download Enable

Copyright

We respect the copyrights and intellectual property rights of all users. All uploaded documents are either original works of the uploader or authorized works of the rightful owners.

  • Not to be reproduced or distributed without explicit permission.
  • Not used for commercial purposes outside of approved use cases.
  • Not used to infringe on the rights of the original creators.
  • If you believe any content infringes your copyright, please contact us immediately.

Support

For help with questions, suggestions, or problems, please contact us