Unterberg-Buchwald et al. Journal of Cardiovascular Magnetic Resonance (2017) 19:45 DOI 10.1186/s12968-017-0357-3 RESEARCH Open Access Targeted endomyocardial biopsy guided by real-time cardiovascular magnetic resonance Christina Unterberg-Buchwald1,3,6*† , Christian Oliver Ritter3,6†, Verena Reupke2, Robin Niklas Wilke3,6, Christine Stadelmann4, Michael Steinmetz3,5, Andreas Schuster1,3, Gerd Hasenfuß1,3, Joachim Lotz3,6 and Martin Uecker3,6 Abstract Background: Endomyocardial biopsies (EMB) are an important diagnostic tool for myocarditis and other infiltrative cardiac diseases. Routinely, biopsies are obtained under fluoroscopic guidance with a substantial radiation burden. Despite procedural success, there is a large sampling error caused by missing the affected myocardium. Therefore, multiple (>6) biopsies are taken in the clinical setting. In cardiovascular magnetic resonance (CMR), late gadolinium enhancement (LGE) depicts areas of affected myocardium in myocarditis or in other infiltrative cardiomyopathies. Thus, targeted biopsy under real-time CMR image guidance might reduce the problem of sampling error. Methods: Seven minipigs of the Goettingen strain underwent radiofrequency ablation in the left ventricle. At least two focal lesions were induced on the lateral wall in five and the apex in two animals. Each ablation lesion was created by two consecutive 30 sec ablations (max. 30 W, temperature 60–64 °C). Biopsies were taken immediately after lesion induction using a commercially available 7 F conventional bioptome under fluoroscopic guidance at the ablation site. Afterwards the animals underwent CMR and lesion visualization by LGE at 3T. The lesions were then targeted and biopsied under CMR-guidance using a MR-conditional bioptome guided by a steerable catheter. Interactive real-time (RT) visualization of the intervention on an in-room monitor was based on radial FLASH with nonlinear inverse reconstruction (NLINV) at a temporal resolution of 42 ms. All samples underwent a standard histological evaluation. (Continued on next page) * Correspondence: [email protected] †Equal contributors 1University Medical Center Goettingen, Clinic of Cardiology and Pneumology, Goettingen, Germany 3DZHK (German Centre for Cardiovascular Research), Partner Site Goettingen, Berlin, Germany 6University Medical Center Goettingen, Institute for Diagnostic and Interventional Radiology, Goettingen, Germany Full list of author information is available at the end of the article © The Author(s). 2017 Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated. Unterberg-Buchwald et al. Journal of Cardiovascular Magnetic Resonance (2017) 19:45 Page 2 of 10 (Continued from previous page) Results: Radiofrequency ablation was successful in all animals. Fluoroscopy-guided biopsies were performed with a success rate of 6/6 minipigs - resulting in a nonlethal pericardial effusion in one animal. Visualization of radiofrequency lesions by CMR was successful in 7/7 minipig, i.e. at least one lesion was clearly visible. Localization and tracking of the catheters and the bioptome using interactive control of the imaging plane was achieved in 6/6 MP; however in the animal with a large pericardial effusion after EMB under fluoroscopy no further EMB was attempted for safety reasons. Biopsies under interactive RT-CMR guidance were successfully performed in 5/6 animals, in one animal the bioptome reached the lesion, however the forceps did not cut out a sample. Specimens obtained under CMR guidance contained part of the lesion in 6/15 (40%) myocardial specimens and in 4/5 (80%) animals in which samples were achieved. Conventional biopsies revealed ablation lesions in 4/17 (23.5%) specimens in 3/6 minipigs (50%). Conclusion: Focal lesions induced by radiofrequency ablation in a minipig model are a useful tool for CMR-guided biopsy studies. In contrast to fluoroscopy, CMR provides excellent visualization of lesions. Interactive real-time CMR allows excellent passive tracking of the instruments and EMB provides significantly superior sampling accuracy compared to fluoroscopy-guided biopsies. Nonetheless, further improvements of MR-compatible bioptomes and guiding catheters are essential before applying this method in a clinical setting. Keywords: Endomyocardial biopsy, CMR, Targeted biopsy, Real-time MRI Background applied in non moving organs like breast [10], liver [11], Endomyocardial biopsies (EMB) are an important diag- kidney [12], prostate [13] or brain [14] using MR- nostic tool in myocarditis, arrhythmias, cardiac tumors, compatible needles. MR-safe cardiac bioptomes are still storage disease, cardiac allograft rejection and other car- not commonly available. Lossnitzer [15] evaluated a pre- diac diseases with unknown origin. With DNA and RNA clinical MR-conditional bioptome in an ex vivo animal detection EMB evolved into an important diagnostic tool heart model. The NIH group [16] recently demonstrated [1, 2] proposed by the AHA and ACC [3] as well as ESC the feasibility of CMR-guided EMB in an in vivo swine [4]. However, there are concerns about the large sam- model with extended infarct scars. However, clinically pling error and the adequate clinical setting for biopsies many cardiac conditions are associated with small circum- is still in debate. Therefore, in today’s clinical routine script lesions rendering targeted biopsies even more chal- multiple (5–6) biopsies are taken from either the left or lenging. Therefore, we developed an animal model right ventricle under fluoroscopic guidance with a radi- with distinct left ventricular lesions created by con- ation burden for the patient and the interventionalist. trolled radiofrequency ablation. The aim of our study Despite procedural success with overall (major and was to show that targeted EMB of focal myocardial minor) complication rates from 0% [5] to 5% [6, 7] miss- lesions is feasible by real-time CMR guidance using a ing the affected myocardium results in a limited diag- clinical 3T scanner, i.e. that continuous real-time nostic value providing a diagnostic result in only 25.5% CMR in three dimensions enables a controlled posi- of clinical cases [6]. Thus, it seems desirable to use a tioning of the guidewire, the guiding catheter, the method for visualization of the diseased parts of the heart bioptome and the biopsy itself. Furthermore, we in order to take directed samples from the affected myo- aimed at showing that targeted biopsy has a lower cardium. Cardiovascular magnetic resonance (CMR) has a sampling error under real-time CMR compared to superior soft-tissue contrast compared to X-ray [8, 9] and fluoroscopically controlled biopsies. allows arbitrary orientation of the imaging plane in three dimensions without exposure to ionizing radiation. Hence, Methods targeted EMB under real-time CMR guidance could solve Animal model the sampling problem by reducing the need for multiple Seven minipigs of the Goettingen strain weighing 30– biopsies. Nevertheless, interventional CMR has to solve 46 kg and aged 12–24 months received diazepam at a several technical challenges. First, MR-safe and suitable dosage of 0.5 mg/kg oral and 2 mg/kg azaperone and guidewires as well as steerable guiding catheters with 10 mg/kg ketamine i.m. as premedication. Anesthesia distal-tip visualization are mandatory to navigate and reli- was induced by 40–60 mg of propofol and 100–150 μg ably reach the affected myocardium. Furthermore, the of fentanyl. Anesthesia was maintained by 3–4Vol%of magnetic field does not allow the use of conventional me- sevoflurane in 50% oxygen and 50% air and 5–10 μg/kg/ tallic bioptomes due to heating, magnetization and h fentanyl. Postoperative analgesia was provided by massive metal artifacts. Until now, MR biopsies are mostly 50 mg/kg metamizole i.v. and 2.2 mg/kg flunixin i.m.. Unterberg-Buchwald et al. Journal of Cardiovascular Magnetic Resonance (2017) 19:45 Page 3 of 10 Blood gases were regularly monitored and ventilation ventriculography or computer tomography - was per- was adjusted to maintain blood gases in the physiologic formed prior to radiofrequency ablation and X-ray range. Surface electrocardiogram, concentration of car- guided endomyocardial biopsy. At least 15 min. post ab- bon dioxide and oxygen level in the blood were moni- lation an 8.5 F deflectable guiding catheter was intro- tored using MR-compatible LCD monitoring system duced into the left ventricle. The guiding catheter (Precess 3160, Invivo, Orlando, Florida, USA). (St.Jude Medical, St. Paul, USA, Fast-Cath, SRO, 8.5 F) All experiments consisted of five parts: was advanced to the ablated myocardium under fluoro- scopic guidance using the same angulations of the radio- 1. radiofrequency ablation in the left ventricle for frequency procedure with help of the stored pictures of creation of circumscript lesions under fluoroscopic the radiofrequency ablation sites as guidance. Endomyo- guidance, cardial biopsies were then taken utilizing a conventional 2. endomyocardial biopsy (aim:
Details
-
File Typepdf
-
Upload Time-
-
Content LanguagesEnglish
-
Upload UserAnonymous/Not logged-in
-
File Pages10 Page
-
File Size-