Electron Tomography—A Tool for Ultrastructural 3D Visualization in Cell Biology and Histology

Electron Tomography—A Tool for Ultrastructural 3D Visualization in Cell Biology and Histology

main topic Wien Med Wochenschr (2018) 168:322–329 https://doi.org/10.1007/s10354-018-0646-y Electron tomography—a tool for ultrastructural 3D visualization in cell biology and histology Josef Neumüller Received: 16 March 2018 / Accepted: 22 June 2018 / Published online: 6 August 2018 © The Author(s) 2018 Summary Electron tomography (ET) was developed lischen Grundlagen, Möglichkeiten und Grenzen. Die to overcome some of the problems associated re- Entwicklung innovativer Verfahren wird besprochen, constructing three-dimensional (3D) images from 2D relevante Untersuchungen aus dem Institut der Au- election microscopy data from ultrathin slices. Virtual toren und in Zusammenarbeit mit anderen werden sections of semithin sample are obtained by incre- dargestellt. Durch die ET erschließt sich die dritte mental rotation of the target and this information is Dimension auf der ultrastrukturellen Ebene, sie stellt used to assemble a 3D image. Herein, we provide einen Meilenstein in der strukturellen Molekularbio- an instruction to ET including the physical principle, logie dar. possibilities, and limitations. We review the develop- ment of innovative methods and highlight important Schlüsselwörter Elektronentomografie · 3D-Darstel- investigations performed inourdepartmentandwith lung · Ultrastruktur · Zellbiologie · Histologie our collaborators. ET has opened up the third di- mension at the ultrastructural level and represents a Introduction milestone in structural molecular biology. When inspecting the organization of organelles of Keywords Electron tomography · 3D visualization · a particular cell or the cellular arrangement in sev- Ultrastructure · Cell biology · Histology eral tissues at the ultrastructural level, we usually start from two-dimensional images. We then inter- Elektronentomografie – Ein Verfahren zur 3D- polate a three-dimensional(3D)imagefromsections Visualisierung von ultrastrukturellen Details in at different heights of the same site of an electron Zellbiologie und Histologie microscopic (EM) preparation. The resulting recon- struction, conceived in our imagination, can lead to Zusammenfassung Die Elektronentomographie (ET) missing details or general misinterpretations. One of wurde entwickelt, um einige der Probleme zu lö- the first successful attempts to overcome these prob- sen, die mit der Rekonstruktion dreidimensionaler lems concerned the 3D reconstruction from ultrathin (3-D-)Bilder aus 2-D-Elektronenmikroskopiedaten serial sections varying between 60 and 80nm. [1–3]. von Ultradünnschnitten einhergehen. Durch die stu- However, such studies are time consuming and re- fenweise erfolgende Rotation des Ziels werden virtuel- quire a skilled technical laboratory staff. Therefore, le Schnitte von Semidünnschnitten gewonnen, diese scientists using electron microscopy sought alterna- Informationen dienen der Zusammenstellung eines tive methods, such as electron tomography (ET). 3-D-Bilds. Im vorliegenden Beitrag geben die Autoren In this review, we give a short introduction into ET eine Anleitung für die ET einschließlich der physika- and a historical overview of its development in rele- vant important centers. We discuss the principle of ET, including possibilities and limitations. We also re- J.Neumüller,Ao.Univ.Prof.Dr.() fer to ET investigations carried out in our department Center of Anatomy and Cell Biology, Department of Cell and Developmental Biology, Medical University of Vienna, and mention by name scientists who were helpful in Schwarzspanierstraße 17, 1090 Vienna, Austria establishing this method in our EM laboratory. We [email protected] also summarize the development of new advanced 322 Electron tomography—a tool for ultrastructural 3D visualization in cell biology and histology K main topic methods, such as focused ion beam scanning tomog- raphy (FIB-SEM), serial block-face scanning electron microscopy (SBF-SEM), axial bright-field scanning transmission electron microscopy (STEM) tomogra- phy, and array tomography (AT). Technical principles of ET, opportunities and limitations ET takes place in a transmission electron microscope (TEM), in a TEM equipped with scanning device (STEM), or in special scanning electron microscopes at an acceleration voltage of at least 200 kV. In ET using TEM, a semithin section mounted on a grid is inserted into a high-tilt holder, which can be tilted by a eucentric goniometer over a range of maximally –70° to +70°. Typically, depending on the specimen, a range of –65° to +65° is used. The thickness of the semithin sections is restricted to 300nm in TEM mode and to 1µm in STEM mode. ET requires a eu- centric goniometer with a complex stage, where the grid holder can be moved in directions of the x-, y-, and z-axes. Computer programs for correcting dis- locations and providing automatic adjustment of the eucentric height and focus (autofocus) during tilting are therefore indispensable. During tilting, digital im- ages are acquired via a sensitive CCD camera at preset intervals (usually 1°). The tilting limits are set by the Fig. 1 Flowchart showing the sequence of electron to- mographic (ET) processing steps as performed with our edge of the grid and the fixation ring at the holder, instrumentation (Tecnai 2G, FEI company, Eindhoven, The which would cover the section at higher tilting angles. Netherlands). Detailed explanations: aGain acquisition: op- This reduces the 3D information and is referred as timizing the image conditions by performing an illumination “missing wedge”. This disadvantage can be partially intensity calibration curve. bHolder calibration: recording of compensated by carrying out a second tomogram dislocations of a region of interest (ROI) during acquisition us- from the same ROI, usually at an angle of 90° to the ing a particular specimen holder (high-tilt or rotation holder). This setting must be performed only once after changing the first axis (dual axis tomography), reducing the miss- c cathode (in our case, a LaB6 cathode). Setting of acquisition ing wedge to a missing cone and providing higher details: range of tilting angle in the+ and – direction, increment 3D information. The digital images are stored in for image acquisition, options for fine tuning (tracking before stack files, which are put through further processing or tracking after taking of an image at a particular tilt angle), software packages [4]. The next step is the recon- choosing whether stack files (*.mrc) or single images should struction, transforming the projections in a Fourier d be stored and into which directory. Import of stack files into space, from which they are placed as virtual slices the IMOD software (Boulder Laboratories for 3D Electron Mi- croscopy of Cells, University of Colorado, USA): The header into a virtual 3D volume. This method is referred of the stack file will read-out by the IMOD software. eCreating as weighted back projection (WBP). Corresponding a seed model by using fiducial markers: appropriate markers structures through the whole 3D space can be traced (gold particles) are checked in the horizontal projection and and connected by the computer in order to obtain a seed model is generated automatically in which respective a 3D model. Beside the WBP method, the simulta- markers of every tilt projection are identified. If this proce- neous iterative reconstruction technique (SIRT; [5]), dure doesn’t work automatically because of bad contrast in and the discrete algebraic reconstruction technique particular projections, the generation of the seed model can also be performed manually. By running a dual-axis ET, the (DART; [6]) have been developed in order to improve position of the fiducial markers from the first axis tilt series the image information. Fig. 1 shows a flowchart of has to be transferred to the second axis tilt series. fCreating the consecutive steps in ET as used in our laboratory. a boundary model allows elimination of image features of no Compared with 3D reconstructions based on serial interest by sampling three regions of the tomogram, ones sections, ET provides significant advantages in z-axis computed from near the top, middle, and bottom of the tilt resolution of the virtual slices which correlates with images. gDrawing a model and creating a movie uses the Amira 5.3.5 software (Mercury Computer Systems, Merignac, the voxel size in nanometers, depending on the im- Cedex, France). WBP, SIRT: reconstruction using Weighted age resolution and the EM magnification used for ET Back Projection or Simulaneous Itinerative Reconstructon acquisition. Technique respectively K Electron tomography—a tool for ultrastructural 3D visualization in cell biology and histology 323 main topic Historical overview and state of the art siastic members together with Margit Pavelka, Josef Neumüller, Claudia Meisslitzer-Ruppitsch, Carmen Investigations using ET started in the last decade of Ranftler, and Monika Vetterlein, and the technicians the past century, as summarized in 1997 [7]. Subse- Peter Auinger, Ivanna Fedorenko, Ulrich Kaindl, Beat- quently, several reviews described the technique [8, rix Mallinger, Thomas Nardelli, and Elfriede Scherzer. 9]. Throughout many years, the complex and dynamic To improve the reliability of EM preparations Golgi apparatus architectures were at the center of our by avoiding chemical artifacts, freezing methods interest. To avoid artifacts due to chemical fixation, we in liquid nitrogen, such as the high-pressure fixa- used high-pressure freezing followed by freeze substi- tion (HPF), were developed. Frozen

View Full Text

Details

  • File Type
    pdf
  • Upload Time
    -
  • Content Languages
    English
  • Upload User
    Anonymous/Not logged-in
  • File Pages
    8 Page
  • File Size
    -

Download

Channel Download Status
Express Download Enable

Copyright

We respect the copyrights and intellectual property rights of all users. All uploaded documents are either original works of the uploader or authorized works of the rightful owners.

  • Not to be reproduced or distributed without explicit permission.
  • Not used for commercial purposes outside of approved use cases.
  • Not used to infringe on the rights of the original creators.
  • If you believe any content infringes your copyright, please contact us immediately.

Support

For help with questions, suggestions, or problems, please contact us