Targeted Metabolomics Method Using IC Orbitrap with High Resolution Mass Spectrometry

Targeted Metabolomics Method Using IC Orbitrap with High Resolution Mass Spectrometry

CUSTOMER APPLICATION NOTE 73265 Targeted metabolomics method using IC Orbitrap with high resolution mass spectrometry Authors: Yoshihiro Izumi, Takeshi Bamba, Medical Institute of Bioregulation, Kyushu University, Research Center for Transomics Medicine, Division of Metabolomics Keywords: Anionic compound, sample- pretreatment, ion chromatography, high resolution mass spectrometer, sugar phosphate Goal Introduction In this application note, the analysis method of ion In the principal metabolic pathways which support life chromatography/Fourier transform mass spectrometry such as the glycolytic pathway, pentose phosphate (IC Thermo Scientific™ Q Exactive™ system) was developed pathway, citric acid cycle, and nucleic acid metabolism, to establish a versatile method of detecting polar metabolites most of the intermediate metabolites are highly polar (particularly anionic metabolites), the isolation and analysis anionic compounds. Therefore, accurate profiling and of which are complicated despite their importance in vivo. comparison of these compounds may greatly contribute A number of different water-soluble compound standards to the understanding of living systems. Various methods were used, and the retention time and MS/MS spectra for have been used previously to measure anionic metabolites. more than 350 types of compounds that can be analyzed CE-MS1,2, ion-pair LC-MS3 and HILIC LC-MS4 are the main with this system were acquired to construct a metabolite approaches, and each has advantages and disadvantages. library. Additionally, using cultured animal cells and human serum samples, optimization of pretreatment for the IC CE-MS has high separation performance, being capable of Q Exactive system was studied. measuring a wide range of anionic metabolites. However, compared to the approach using LC, the measurement throughput time and reproducibility of electrophoresis are lower. For these reasons, CE-MS requires acquisition of high level skills for stable operation5. In addition, because the sample loading volumes should be 30 nL or less, ingenuity is needed for pretreatment. On the other hand, ion-pair LC-MS has relatively higher IC Q Exactive system measurement conditions sensitivity without requiring special techniques, enabling A) Ion Chromatography (IC) conditions high throughput analysis. However, the effects of ion- Equipment Thermo Scientific™ Dionex™ ICS-5000+ HPIC pairing reagents on a mass spectrometer cannot be system ignored, and it is essential to use a mass spectrometer IC analytical Dionex IonPac AS11-HC-4µm, 2 × 250 mm specialized for ion-pairing reagents. HILIC LC-MS is an column (P/N 4078035) approach without ion-pairing reagents and so does not IC guard Dionex IonPac AS11HC-9µm, 2 × 50 mm require a specialized machine, but, due to insufficient column (P/N 078036) ™ ™ ™ separation of many existing isomers, the range of Trap column Thermo Scientific Dionex CR-ATC 500 Trap column (P/N 075550) measurable anionic metabolites becomes narrower. Column 30 °C Due to the above, it is desirable to develop an analytical temperature approach that has a resolving power capable of separating Eluent KOH, Thermo Scientific™ Dionex™ EGC 500 and detecting a wider range of anions, high sensitivity to KOH Potassium Hydroxide Eluent Generator detect trace amounts of metabolites that have physiological Cartridge (P/N 075778) activities, a throughput capable of acquiring data from Gradient 10 mM (0 min) – 100 mM (24 min) – 100 mM various biological samples, and is rugged. (27 min) – 10 mM (27.1 min) – 10 mM (30 min) Suppressor Thermo Scientific™ Dionex™ AERS™ 500e Electrolytic Suppressor, 2 mm, 75 mA, In this application note, IC was used for the separation external water mode (P/N 302662) part. IC was performed with an anion-exchange column Flow rate 0.3 mL/min (Thermo Scientific™ Dionex™ IonPac AS11-HC-4µm) and Make-up 0.1 mL/min (MeOH with 1 mM ammonium non-volatile potassium hydroxide as the eluent for IC under acetate) a concentration gradient (10–100 mM). Generally, salt Injection volume 2 μL contamination is a concern when IC is connected directly B) Q Exactive mass spectrometer conditions to MS. However, using an electrolytically regenerated Equipment Thermo Scientific Q Exactive mass suppressor that removes nonvolatile salt to make the eluent spectrometer water enables routine analysis with the IC-MS system. A Ionization Positive/Negative ESI high-resolution and accurate mass (HRAM) Q Exactive method mass spectrometer was used as a detector in order to Spray voltage 3.5 kV / -2.0 kV acquire a clear and highly accurate profile. Vaporizer 400 °C temperature Pump Sheath gas 60 arb Aux gas 10 arb Ultrapure water Ion transfer tube 250 °C temperature Suppressor Conductivity S-Lens level 50 (salt removal) detector [Full MS] Resolution 70,000 Dionex ICS-5000+ Waste HPIC system Mass width m/z 70–1050 [Data dependent-MS2] Pump Resolution 17,50 0 Enhanced TopN 5 ionization Isolation m/z 1.5 Q Exactive window Methanol Mass Spectrometer Intensity 5,000 threshold Figure 1. IC Q Exactive system Normalized 10, 30, 45 collision energy 2 Sample preparation Checking for measurable metabolites by IC The standard products used to construct an IC Q Exactive In-house authentic standards (hydrophilic metabolites): 600 metabolite library were prepared in ultrapure water to a final Acquisition of MS and MS/MS spectra with concentration of 50 μM. HeLa cells cultured in a 10 cm high-precision (mass resolution: 70,000, mass accuracy: < 5 ppm) plate (1×10 6–1×107) were washed twice with 10 mL of each wash solvent (ultrapure water, 150 mM ammonium acetate Storage of data (RT, MS and MS/MS spectra) solution, 150 mM ammonium carbonate solution, 150 mM PBS, all at 4°C) after the culture medium in the plate was removed. Subsequently, the metabolism was quenched by the addition of 1 mL cold methanol (-30°C), and the cells Development of a wide targeted and user-friendly were scraped. The resulting cell suspension was collected metabolomics analytical system by IC Q Exactive in a 2 mL tube. Subsequently, 400 μL chloroform was added and extraction was performed. After centrifugation, Figure 2. IC Q Exactive development workflow 700 μL supernatant was collected. After ultrapure water (400 μL) and chloroform (300 μL) were added, two-phase Library Manager 2.0 Thermo Scientic™ TraceFinder™ Software, fractionation was performed by centrifugation, and the version 3.1 upper phase (water/methanol) was collected. Then, the sample was completely dried with a centrifugal evaporator. The sample was then re-dissolved in 50 μL ultrapure water and placed on the analytical system. For 20 μL serum, 1 mL mixed solvent (methanol : In-house chloroform : water = 2.5 : 1 : 1) was added, and extraction database was performed as done with the cell samples. After two- Positive-ion 264 phase fractionation, the upper phase was collected. After Negative-ion 336 that, the fraction of polar proteins was removed using a Total Compounds 359 5,000 molecular weight cutoff ultrafiltration membrane. The Figure 3 . IC Q Exactive metabolite library resulting filtrate was dried at reduced pressure, and re- dissolved in 50 μL ultrapure water to be used as a sample. Assessment of IC separation performance in targeted Results and discussion metabolome analysis Construction of IC Q Exactive metabolite library Among 359 types of metabolites successfully separated Using the IC Q Exactive system, the conditions for IC and analyzed using IC Q Exactive, coenzymes such as separation were optimized, and MS and MS/MS spectral ATP were included, which were difficult to analyze by the libraries were constructed (Figure 2). First, the separation conventional approaches such as CE-MS, ion-pair LC-MS conditions for IC analysis were studied using 600 types of and HILIC LC-MS. hydrophilic metabolite standard products that are anionic, cationic, or amphoteric. As a result, 264 types in positive These metabolites are highly polar compounds containing ion mode and 336 types in negative ion mode, a total of two or more phosphate groups (logP < –6) and strong 359 types of precursor ions, were confirmed (common: metal chelators. Therefore, it is suggested that, in the 241 types, positive ion only: 23 types, negative ion only: conventional approaches, absorption of compounds 95 types). Subsequently, MS/MS spectra were acquired with phosphate groups onto the wetted metal part of the for the confirmed precursor ions, and their information was apparatus such as stainless steel tubing occurs, leading to stored in Library Manager 2.0 (Figure 3). peak tailing and reduced sensitivity and reproducibility. 3 On the other hand, it is now clear that the IC Q Exactive approach using IC Q Exactive has high reproducibility with system enables separation and analysis of such cell extract samples as well, and also higher separation coenzymes, keeping sharp peak shapes (Figure 4). It performance compared to the conventional approaches is presumably because coenzymes, which are highly (Figure 4). Furthermore, because trace amounts of polar anionic metabolites, are suitable for the separation secondary messengers such as cAMP were also detected characteristics of IC, “separation by interaction between ion from cell extracts, this approach can be considered as a exchangers and ion types (ion valency, ionic radius and ion highly sensitive analysis system (Figure 4). hydrophobicity”,4 and also adsorption of compounds with phosphate groups onto metals, do not occur as all flow Most of the metabolites measured using this analysis paths in

View Full Text

Details

  • File Type
    pdf
  • Upload Time
    -
  • Content Languages
    English
  • Upload User
    Anonymous/Not logged-in
  • File Pages
    6 Page
  • File Size
    -

Download

Channel Download Status
Express Download Enable

Copyright

We respect the copyrights and intellectual property rights of all users. All uploaded documents are either original works of the uploader or authorized works of the rightful owners.

  • Not to be reproduced or distributed without explicit permission.
  • Not used for commercial purposes outside of approved use cases.
  • Not used to infringe on the rights of the original creators.
  • If you believe any content infringes your copyright, please contact us immediately.

Support

For help with questions, suggestions, or problems, please contact us