<p> ABB Instrumentation Division Inc. Warminster, Pa.</p><p>Magnetic Flowmeter Calibration Test Report</p><p>Sales Order No.:284090 Date :29 AUG 2000 Serial No.:00W035759 Line Item:1 Meter Size :8 inch(200mm) Model No.:MFF201361111004ER Sec. Serial No.:00W035760 Sec Model No.:MFE4ER140111 Max Flow: 600.000 GPM Sp. Gr. : 1.000</p><p>Sensor Factor 1 :1.4934 Sensor Factor 2 : -5.0 Sensor Factor 3 : 5 Sensor Factor 4 :1.0000</p><p>Run # Actual GPM Indicated GPM Error % RATE 01 97.473 97.365 -.111 02 329.441 329.691 +.076 03 590.859 591.528 +.113</p><p>All Flowmeters are hydraulically calibrated in accordance with ANSI/Z540 and are traceable to the NATIONAL INSTITUTE OF STANDARDS AND TECHNOLOGY. The ABB Instrumentation Div. Warminster facility is certified to ISO 9001.</p><p>This Calibration report may not be reproduced, except in full, without written permission.</p><p>Hydraulic test performed by :G. Myers</p><p>Required Accuracy :0.15% of Rate</p><p>TEST EQUIPMENT : QC # Description E1934 TIMER E3222 DVM M**85 SCALE M**84 SCALE E3544 COUNTER </p><p>NEWPORT6.BAS, RAM 120998 11:14:47 </p>