European Heart Journal (1998) 19, 1879–1888 Article No. hj981199 The transthoracic impedance cardiogram is a potential haemodynamic sensor for an automated external defibrillator P. W. Johnston*, Z. Imam†, G. Dempsey†, J. Anderson† and A. A. J. Adgey *Regional Medical Cardiology Centre, Royal Victoria Hospital, Belfast; †Northern Ireland Bioengineering Centre, Belfast, Northern Ireland Aims The American Heart Association has endorsed the agonal rhythm (20), electromechanical dissociation (22), concept of Public Access Defibrillation. However, there ventricular tachycardia (27) and sinus rhythm (5). These have been reports of inappropriate direct current shocks rhythms were divided into those associated with haemo- from automatic external defibrillators. The specificity of dynamic collapse i.e. no pulse — asystole, ventricular fibril- automatic external defibrillators for shockable rhythms lation, agonal rhythm, electromechanical dissociation and may be improved by the incorporation of a haemodynamic shockable ventricular tachycardia (associated with loss of sensor. consciousness, pulselessness or a systolic blood pressure of <80 mmHg) (Group 1) and those associated with a satis- Methods and Results This study examined the use of four factory cardiac output i.e. non-shockable ventricular tachy- parameters extracted from the impedance cardiogram i.e. cardia (conscious with a pulse) and sinus rhythm (Group 2). Peak dz/dt (the peak of the impedance cardiogram On univariate analysis each of the four impedance cardio- measured from the line dz/dt=0 Ùs"1), Peak-trough (the gram parameters were significantly greater in Group 2 than peak-to-trough measurement of the impedance cardiogram Group 1 (P<0·001). On multivariate analysis the par- Ùs"1), Area 1 (the area under the C wave of the impedance ameters which best differentiated the two groups were Area cardiogram above the line dz/dt=0 mÙ) and Area 2 (the 1 and Peak-trough. area under the impedance cardiogram 50 ms on either side of the Peak and above the line dz/dt=0 mÙ) as predictors Conclusion Thus the impedance cardiogram is a of cardiac output. potential haemodynamic sensor for an automatic external defibrillator. At 116 cardiac arrest calls the ECG and impedance cardio- (Eur Heart J 1998; 19: 1879–1888) gram were recorded through two ECG/defibrillator pads placed in an antero-apical position. Nine recordings were Key Words: Impedance cardiogram, automatic external rejected for artefact. The rhythm recorded in the remaining defibrillator, cardiac arrest. 107 calls was asystole (19), ventricular fibrillation (14), Introduction survivors from sudden cardiac death and by lay persons in large public gatherings[3,4]. The American Heart Survival from out-of-hospital ventricular fibrillation has Association has endorsed the concept of Public Access [5] been possible since the development of the portable Defibrillation . However, there have already been defibrillator[1]. The most important factor influencing reports of the inappropriate delivery of direct survival is the delay to defibrillation[2]. The scope of current shocks to patients without ventricular [6–8] early defibrillation has been extended by the develop- tachyarrhythmias . This has led the U.S. Food and ment of the automatic external defibrillator. This device Drug Administration to issue a safety alert with one of [8] has been used successfully by family members of these devices . The use of automatic external defibrilla- tors by minimally trained individuals who lack the skills to differentiate cardiac arrest from other causes of Revision submitted 9 June 1998, and accepted 10 June 1998. collapse will increase the potential for inappropriate Correspondence: Prof A. A. J. Adgey, Regional Medical defibrillation. Clearly, the inclusion of a non-ECG Cardiology Centre, Royal Victoria Hospital, Belfast BT12 6BA, haemodynamic sensor in an automatic external Northern Ireland. defibrillator device could increase its specificity. 0195-668X/98/121879+10 $18.00/0 1998 The European Society of Cardiology 1880 P. W. Johnston et al. Impedance cardiography is a non-invasive method for measuring cardiac output, and parameters derived from the impedance cardiogram have been used as indices of myocardial contractility and aortic blood flow[9–18]. The impedance of the thorax (Z) can be recorded by passing a high frequency low amplitude current between two electrodes and recording the result- ing voltage. Ejection of blood into the aorta causes small fluctuations in this impedance (ÄZ). The impedance cardiogram is a recording of the first time derivative of ÄZ (i.e. dz/dt) against time[19]. Traditionally the impedance cardiogram has been recorded using four circumferential band elec- trodes[19]. The upper voltage electrode is placed around the base of the neck and the lower voltage electrode Figure 1 The ECG/impedance cardiogram unit (A) around the thorax at the level of the xiphisternum. The connected to a portable defibrillator (B) and two ECG/ two outer current electrodes are placed at least 3 cm defibrillator pads (C). away from the voltage electrodes. Clearly it was not practical to employ such an electrode configuration in the cardiac arrest setting. Thus, a system was developed prototype device. This device could use either the four- in which the impedance cardiogram could be recorded electrode technique or the new two ECG/defibrillator through two ECG/defibrillator pads, one placed at the pad technique. It passed a high frequency (64 kHz) low second right intercostal space to the right of the sternum amplitude constant alternative current (1 mA RMS) be- just beneath the clavicle and the other placed over the tween two electrodes (i.e. the outer two band electrodes fifth left intercostal space in the mid-clavicular line. in the four electrode technique and the two ECG/ The purposes of this research were firstly, to defibrillator pads in the two electrode technique) and compare the impedance cardiogram recordings using a recorded the impedance cardiogram through two elec- traditional four band electrode technique with the novel trodes (i.e. the inner two band electrodes in the four two ECG/defibrillator pad technique and secondly, to electrode technique or the two ECG/defibrillator pads in determine if one or more of the impedance cardiogram the two electrode technique). The signals were digitized parameters, recorded through two ECG/defibrillator and stored on a portable computer for subsequent pads, could act as a haemodynamic sensor for an analysis using a commercial software package automated external defibrillator. (Snap-Master 2·00, HEM Data Corp). The circuitry for the two ECG/defibrillator pad Methods technique was then incorporated into a portable ECG/ impedance cardiogram recording unit and hardware was Controls and patients added to protect the device from high voltage direct current shocks. This unit could be connected to each of The impedance cardiogram was recorded in 20 male four portable defibrillators (Liteguard 6, Temtech, subjects (mean (&SD) age, 26·1 (&5·8)) in sinus Bangor, N.I.) using a locking plug (Hewlett Packard) rhythm using sequentially the traditional four-band (Fig. 1). Again this unit passed a high frequency electrodes and the new two ECG/defibrillator pad (64 kHz) low amplitude constant alternative current technique. (1 mA RMS) between the two ECG/defibrillator pads. Over a 14-month period, simultaneous record- The ECG and impedance cardiogram were detected and ings of the ECG and impedance cardiogram were made the signals simultaneously digitized and stored on at 116 cardiac arrest calls in 110 patients. The baseline memory cards for analysis off-line. characteristics age, sex, site of arrest, delay to cardio- pulmonary resuscitation and delay to intensive care were documented for each of the patients. There were 39 Procedure females and 71 males. The age of one out-of-hospital arrest patient was not known. Of the remaining 109 Immediately on arrival at a cardiac arrest patient, car- patients the mean age was 66 years (range 38–87 years). diopulmonary resuscitation continued or was initiated Sixty cardiac (51·7%) arrest calls were attended by the by the junior doctor manning the cardiac ambulance or mobile coronary care unit outside hospital and 56 attending the in-hospital cardiac arrest and ECG/ (48·3%) were inside the hospital. defibrillator pads positioned as for cardiac arrest management. To determine the cardiac rhythm, cardio- Equipment pulmonary resuscitation was momentarily discontinued and a 10 s recording of the ECG and impedance cardio- In the control subjects the ECG, impedance cardiogram gram, free of cardiopulmonary resuscitation artefact, and baseline impedance (Z) were recorded using a was made. Eur Heart J, Vol. 19, December 1998 The transthoracic impedance cardiogram 1881 2·5 mV 1 mV 0 ECG –1 mV –2·5 mV C 1·5 Ω s–1 B dz/dt O A 0 ICG X Y –1·5 Ω s–1 0 0·5 1·0 Time (s) Figure 2 Features of the impedance cardiogram on a simultaneous recording of the ECG and impedance cardiogram at 25 mms"1 during sinus rhythm. A=A wave representing atrial systole, B=B point corresponding with aortic valve opening, C=C wave represent- ing ventricular systole, X=X point corresponding with aortic valve closure, Y=Y point corresponding with pulmonary valve closure and O=O wave representing ventricular filling. ICG=impedance cardiogram. Impedance cardiogram waveform plex. In the cases of ventricular fibrillation no R waves were present and ensemble averaging was performed The impedance cardiogram waveform has A, C and O using the peaks of the fibrillatory waveform that were waves and contains B, X and Y points (Fig. 2). The A greater than a threshold value (0·2 mV) as reference and C waves occur during atrial and ventricular systole, points. Similar amplitude criteria are employed in the respectively, and the O wave corresponds with ventricu- detection algorithms of current automatic external defi- lar filling. The B point has been related to aortic brillators[21]. In agonal rhythm, the peak of the ECG valve opening and the X and Y points to aortic and complex was used as a reference point.
Details
-
File Typepdf
-
Upload Time-
-
Content LanguagesEnglish
-
Upload UserAnonymous/Not logged-in
-
File Pages10 Page
-
File Size-