Lecture 4 Determination of Protein Concentration by Ultraviolet Spectroscopy
Total Page:16
File Type:pdf, Size:1020Kb
Load more
Recommended publications
-
Developing Back Reflectance Absorbance As a Useful Technique
Design of a Simple Cryogenic System for Ultraviolet-Visible Absorption Spectroscopy with a Back-reflectance Fiber Optic Probe Andrew Vinyard, Kaj Hansen, Ross Byrd, Douglas A Stuart * and John Hansen * Department of Chemistry, University of West Georgia *Corresponding Authors, email: [email protected] and [email protected] Abstract We report a convenient and inexpensive technique for the rapid acquisition of absorption spectra from small samples at cryogenic temperatures using a home built cryostat with novel collection optics. A cylindrical copper block was constructed with a coaxial bore to hold a 4.00 mm diameter EPR tube and mounted on a copper feed in thermal contact with liquid nitrogen. A 6.35 mm diameter hole was bored into the side of the cylinder so a fiber optic cable bundle could be positioned orthogonally to the EPR tube. The light passing through the sample is reflected off of the opposing surfaces of the EPR tube and surrounding copper, back through the sample. The emergent light is then collected by the fiber optic bundle, and analyzed by a dispersive spectrometer. Absorption spectra for KMnO4 were measured between 400 nm and 700 nm. Absorption intensity at 506 nm, 525 nm, 545 nm and 567 nm was found to be proportional to concentration, displaying Beer’s law like behavior. The EPR tube had an internal diameter of 3.2 mm; the double pass of the probe beam through the sample affords a central path length of about 6.4 mm. Comparing these measurements with those recorded on a conventional tabletop spectrometer using a cuvette with a 10.00 mm path length, we consistently found a ratio between intensities of 0.58 rather than the anticipated 0.64. -
2 in 1 Fluorescence and Absorbance Spectrometer
2 in 1 Fluorescence and Application Fluorescence Absorbance Spectrometer: Note How does it work? Life Sciences Figure 1: Duetta (left) and the inside of the Duetta sample compartment (right) with direction of the light path Introduction Duetta™ is a 2-in-1 fluorescence and absorbance excitation energy and the spectrum at longer wavelengths spectrometer from HORIBA Scientific. There are several is acquired to measure the distribution of intensity and benefits of having two different spectroscopies in one energy (wavelength) of the photons emitted. instrument. For high concentration solutions of interest, both primary and secondary inner-filter effects will affect Fluorescence Excitation Spectrum the fluorescence spectrum measured on a standard Acquiring the intensity of photons emitted at a single fluorometer. Having absorbance and fluorescence emission wavelength and scanning the excitation spectroscopy on the same instrument enables Duetta monochromator to excite the population of molecules in and EzSpec software to apply corrections for inner-filter the sample with different wavelengths. A fluorescence effect and provide more accurate data, for a wider range of excitation spectrum is analogous to an absorbance sample concentrations. Another benefit of the two in one spectrum, but is specific to a single emitting species/ instrument is that absorbance results and fluorescence wavelength as opposed to collecting all absorbing species results contain less error since the sample does not move in a sample or solution. from one instrument to another to get both measurements. Standard fluorescence methods such as Emission %Transmittance Spectrum Spectrum and Excitation Spectrum are available on Duetta, This is a ratio, in terms of percentage, of the intensity of but EzSpec software gives a user the ability to measure transmitted light through an absorbing sample (I) compared the Absorbance Spectrum as a stand-alone method or to the transmitted light through a blank solvent (I0). -
Protein Unfolding (Model Peptide Helix/Helix-Coil Transition/Denaturant Effects) J
Proc. Natl. Acad. Sci. USA Vol. 92, pp. 185-189, January 1995 Biochemistry Urea unfolding of peptide helices as a model for interpreting protein unfolding (model peptide helix/helix-coil transition/denaturant effects) J. MARTIN SCHOLTZ*tt, DOUG BARRICK*§, EUNICE J. YORK0, JOHN M. STEWART$, AND ROBERT L. BALDWIN*: *Department of Biochemistry, Stanford University School of Medicine, Stanford, CA 97305-5307; tDepartment of Medical Biochemistry and Genetics, Center for Macromolecular Design, Texas A&M University, College Station, TX 77843-1114; and 1Department of Biochemistry, University of Colorado Health Sciences Center, Denver, CO 80262 Contributed by Robert L. Baldwin, August 29, 1994 ABSTRACT To provide a model system for understand- The peptides examined here consist primarily of alanine;"1 ing how the unfolding of protein a-helices by urea contributes thus side-chain interactions are minimal, and the primary to protein denaturation, urea unfolding was measured for a structural and thermodynamic changes during helix unfolding homologous series of helical peptides with the repeating should result from the peptide backbone. Because these helical sequence Ala-Glu-Ala-Ala-Lys-Ala and chain lengths varying peptides lack hydrophobic cores and long-range tertiary in- from 14 to 50 residues. The dependence of the helix propa- teractions, the thermodynamics of solvent denaturation of gation parameter of the Zimm-Bragg model for helix-coil secondary structure can be measured directly. Although sec- transition theory (s) on urea molarity ([urea]) was deter- ondary structure formation must be an integral part of the mined at 0°C with data for the entire set of peptides, and a protein folding reaction, these other factors (hydrophobic core linear dependence ofIns on [urea] was found. -
Optical and Electron Paramagnetic Resonance Characterization of Point Defects in Semiconductors
Air Force Institute of Technology AFIT Scholar Theses and Dissertations Student Graduate Works 3-1-2019 Optical and Electron Paramagnetic Resonance Characterization of Point Defects in Semiconductors Elizabeth M. Scherrer Follow this and additional works at: https://scholar.afit.edu/etd Part of the Electromagnetics and Photonics Commons Recommended Citation Scherrer, Elizabeth M., "Optical and Electron Paramagnetic Resonance Characterization of Point Defects in Semiconductors" (2019). Theses and Dissertations. 2463. https://scholar.afit.edu/etd/2463 This Dissertation is brought to you for free and open access by the Student Graduate Works at AFIT Scholar. It has been accepted for inclusion in Theses and Dissertations by an authorized administrator of AFIT Scholar. For more information, please contact [email protected]. OPTICAL AND ELECTRON PARAMAGNETIC RESONANCE CHARACTERIZATION OF POINT DEFECTS IN SEMICONDUCTORS DISSERTATION Elizabeth M. Scherrer, Captain, USAF AFIT-ENP-DS-19-M-091 DEPARTMENT OF THE AIR FORCE AIR UNIVERSITY AIR FORCE INSTITUTE OF TECHNOLOGY Wright-Patterson Air Force Base, Ohio DISTRIBUTION STATEMENT A APPROVED FOR PUBLIC RELEASE; DISTRIBUTION UNLIMITED. The views expressed in this thesis are those of the author and do not reflect the official policy or position of the United States Air Force, Department of Defense, or the United States Government. This material is declared a work of the U.S. Government and is not subject to copyright protection in the United States. AFIT-ENP-DS-19-M-091 OPTICAL AND ELECTRON PARAMAGNETIC RESONANCE CHARACTERIZATION OF POINT DEFECTS IN SEMICONDUCTORS DISSERTATION Presented to the Faculty Department of Engineering Physics Graduate School of Engineering and Management Air Force Institute of Technology Air University Air Education and Training Command In Partial Fulfillment of the Requirements for the Doctor of Philosophy Degree Elizabeth M. -
Spectrophotometry Light and Spectra
Page: 1 Spectrophotometry Spectrophotometry and colorimetry are conventional techniques for quantitatively determining substances encountered in biochemistry. All substances in solution absorb light of some wavelength and transmit light of other wavelengths. Absorbance is a characteristic of a substance just like melting point, boiling point, density and solubility. Absorbance can be related to the amount of the substance in solution, thus it can be used to quantitatively determine the amount of substance that is present. Light and Spectra Light or electromagnetic radiation is composed of photons moving in a wave that oscillates along the path of motion. The wavelength of light is defined as the distance between adjacent peaks in the wave and can be further defined by the equation: λ = c / ν where λ is the wavelength, c is the speed of light, and ϖ is the frequency or number of waves passing a certain point per unit time. Photons of different wavelengths have different energies that are given by: E = hc / λ = h ν where h is Planck's constant. Thus, the shorter the wavelength, the greater the energy. Electromagnetic radiation can be divided into various regions according to wavelength: the ultraviolet region has wavelengths 200-400 nm and the visible region has wavelengths of 400-700 nm. There are other regions such as infrared, radio wave, microwave and more, but you will not apply them in this course. In the visible region, lights of different wavelengths have different colors: violet and blue in the low wavelength region and orange and red in the high wavelength region. When a substance in solution appears blue, it means that the substance is absorbing red light and transmitting blue light. -
State Folding of Globular Proteins
Preprints (www.preprints.org) | NOT PEER-REVIEWED | Posted: 20 February 2020 doi:10.20944/preprints202002.0290.v1 Peer-reviewed version available at Biomolecules 2020, 10, 407; doi:10.3390/biom10030407 Review The Molten Globule, and Two-State vs. Non-Two- State Folding of Globular Proteins Kunihiro Kuwajima,* Department of Physics, School of Science, the University of Tokyo, 7-3-1 Hongo, Bunkyo-ku, Tokyo 113-0033, Japan, and School of Computational Sciences, Korea Institute for Advanced Study (KIAS), Seoul 02455, Korea ; [email protected] * Correspondence: [email protected] Abstract: From experimental studies of protein folding, it is now clear that there are two types of folding behavior, i.e., two-state folding and non-two-state folding, and understanding the relationships between these apparently different folding behaviors is essential for fully elucidating the molecular mechanisms of protein folding. This article describes how the presence of the two types of folding behavior has been confirmed experimentally, and discusses the relationships between the two-state and the non-two-state folding reactions, on the basis of available data on the correlations of the folding rate constant with various structure-based properties, which are determined primarily by the backbone topology of proteins. Finally, a two-stage hierarchical model is proposed as a general mechanism of protein folding. In this model, protein folding occurs in a hierarchical manner, reflecting the hierarchy of the native three-dimensional structure, as embodied in the case of non-two-state folding with an accumulation of the molten globule state as a folding intermediate. The two-state folding is thus merely a simplified version of the hierarchical folding caused either by an alteration in the rate-limiting step of folding or by destabilization of the intermediate. -
Electron Paramagnetic Resonance of Radicals and Metal Complexes. 2. International Conference of the Polish EPR Association. Wars
! U t S - PL — voZ, PL9700944 Warsaw, 9-13 September 1996 ELECTRON PARAMAGNETIC RESONANCE OF RADICALS AND METAL COMPLEXES 2nd International Conference of the Polish EPR Association INSTITUTE OF NUCLEAR CHEMISTRY AND TECHNOLOGY UNIVERSITY OF WARSAW VGL 2 8 Hi 1 2 ORGANIZING COMMITTEE Institute of Nuclear Chemistry and Technology Prof. Andrzej G. Chmielewski, Ph.D., D.Sc. Assoc. Prof. Hanna B. Ambroz, Ph.D., D.Sc. Assoc. Prof. Jacek Michalik, Ph.D., D.Sc. Dr Zbigniew Zimek University of Warsaw Prof. Zbigniew Kqcki, Ph.D., D.Sc. ADDRESS OF ORGANIZING COMMITTEE Institute of Nuclear Chemistry and Technology, Dorodna 16,03-195 Warsaw, Poland phone: (0-4822) 11 23 47; telex: 813027 ichtj pi; fax: (0-4822) 11 15 32; e-mail: [email protected] .waw.pl Abstracts are published in the form as received from the Authors SPONSORS The organizers would like to thank the following sponsors for their financial support: » State Committee of Scientific Research » Stiftung fur Deutsch-Polnische Zusammenarbeit » National Atomic Energy Agency, Warsaw, Poland » Committee of Chemistry, Polish Academy of Sciences, Warsaw, Poland » Committee of Physics, Polish Academy of Sciences, Poznan, Poland » The British Council, Warsaw, Poland » CIECH S.A. » ELEKTRIM S.A. » Broker Analytische Messtechnik, Div. ESR/MINISPEC, Germany 3 CONTENTS CONFERENCE PROGRAM 9 LECTURES 15 RADICALS IN DNA AS SEEN BY ESR SPECTROSCOPY M.C.R. Symons 17 ELECTRON AND HOLE TRANSFER WITHIN DNA AND ITS HYDRATION LAYER M.D. Sevilla, D. Becker, Y. Razskazovskii 18 MODELS FOR PHOTOSYNTHETIC REACTION CENTER: STEADY STATE AND TIME RESOLVED EPR SPECTROSCOPY H. Kurreck, G. Eiger, M. Fuhs, A Wiehe, J. -
UV-Vis Spectroscopy: a New Approach for Assessing the Color Index of Transformer Insulating Oil
sensors Article UV-Vis Spectroscopy: A New Approach for Assessing the Color Index of Transformer Insulating Oil Yang Sing Leong 1 ID , Pin Jern Ker 1,*, M. Z. Jamaludin 1, Saifuddin M. Nomanbhay 1, Aiman Ismail 1, Fairuz Abdullah 1, Hui Mun Looe 2 and Chin Kim Lo 2 1 Institute of Power Engineering, College of Engineering, Universiti Tenaga Nasional, Kajang 43000, Selangor, Malaysia; [email protected] (Y.S.L.); [email protected] (M.Z.J.); [email protected] (S.M.N.); [email protected] (A.I.); [email protected] (F.A.) 2 Tenaga Nasional Berhad (TNB) Research Sdn. Bhd., Bandar Baru Bangi, Kajang 43000, Selangor, Malaysia; [email protected] (H.M.L.); [email protected] (C.K.L.) * Correspondence: [email protected] Received: 30 April 2018; Accepted: 15 June 2018; Published: 6 July 2018 Abstract: Monitoring the condition of transformer oil is considered to be one of the preventive maintenance measures and it is very critical in ensuring the safety as well as optimal performance of the equipment. Various oil properties and contents in oil can be monitored such as acidity, furanic compounds and color. The current method is used to determine the color index (CI) of transformer oil produces an error of 0.5 in measurement, has high risk of human handling error, additional expense such as sampling and transportations, and limited samples can be measured per day due to safety and health reasons. Therefore, this work proposes the determination of CI of transformer oil using ultraviolet-to-visible (UV-Vis) spectroscopy. -
Chemically Denatured Structures of Porcine Pepsin Using Small-Angle X-Ray Scattering
polymers Article Chemically Denatured Structures of Porcine Pepsin using Small-Angle X-ray Scattering Yecheol Rho 1, Jun Ha Kim 2, Byoungseok Min 2 and Kyeong Sik Jin 2,* 1 Chemical Analysis Center, Korea Research Institute of Chemical Technology, 141, Gajeong-ro, Yuseong-gu, Daejeon 34114, Korea; [email protected] 2 Pohang Accelerator Laboratory, Pohang University of Science and Technology, 80 Jigokro-127-beongil, Nam-Gu, Pohang, Kyungbuk 37673, Korea; [email protected] (J.H.K.); [email protected] (B.M.) * Correspondence: [email protected]; Tel.: +82-54-279-1573; Fax: +82-54-279-1599 Received: 6 November 2019; Accepted: 12 December 2019; Published: 14 December 2019 Abstract: Porcine pepsin is a gastric aspartic proteinase that reportedly plays a pivotal role in the digestive process of many vertebrates. We have investigated the three-dimensional (3D) structure and conformational transition of porcine pepsin in solution over a wide range of denaturant urea concentrations (0–10 M) using Raman spectroscopy and small-angle X-ray scattering. Furthermore, 3D GASBOR ab initio structural models, which provide an adequate conformational description of pepsin under varying denatured conditions, were successfully constructed. It was shown that pepsin molecules retain native conformation at 0–5 M urea, undergo partial denaturation at 6 M urea, and display a strongly unfolded conformation at 7–10 M urea. According to the resulting GASBOR solution models, we identified an intermediate pepsin conformation that was dominant during the early stage of denaturation. We believe that the structural evidence presented here provides useful insights into the relationship between enzymatic activity and conformation of porcine pepsin at different states of denaturation. -
UV-VIS Nomenclature and Units
IN804 Info Note 804: UV-VIS Nomenclature and Units Ultraviolet-visible spectroscopy or ultraviolet-visible spectrophotometry (UV/VIS) involves the spectroscopy of photons in the UV-visible region. It uses light in the visible and adjacent near ultraviolet (UV) and near infrared (NIR) ranges. In this region of the electromagnetic spectrum, molecules undergo electronic transitions. This technique is complementary to fluorescence spectroscopy, in that fluorescence deals with transitions from the excited state to the ground state, while absorption measures transitions from the ground state to the excited state. UV/VIS is based on absorbance. In spectroscopy, the absorbance A is defined as: (1) where I is the intensity of light at a specified wavelength λ that has passed through a sample (transmitted light intensity) and I is the intensity of the light before it enters the sample or 0 incident light intensity. Absorbance measurements are often carried out in analytical chemistry, since the absorbance of a sample is proportional to the thickness of the sample and the concentration of the absorbing species in the sample, in contrast to the transmittance I / I of a 0 s ample, which varies exponentially with thickness and concentration. The Beer-Lambert law is used for concentration determination. The term absorption refers to the physical process of absorbing light, while absorbance refers to the mathematical quantity. Also, absorbance does not always measure absorption: if a given sample is, for example, a dispersion, part of the incident light will in fact be scattered by the dispersed particles, and not really absorbed. Absorbance only contemplates the ratio of transmitted light over incident light, not the mechanism by which light intensity decreases. -
Quantifying Protein Concentration Using UV Absorbance Measured by the Ocean HDX Spectrometer by Yvette Mattley, Ph.D
Quantifying Protein Concentration using UV Absorbance Measured by the Ocean HDX Spectrometer by Yvette Mattley, Ph.D. Application Note KEYWORDS Proteins are a common sample in clinical, diagnostic and re- • Protein search labs. Studies involving protein extraction, purification or • Bovine serum albumin labeling, or working with proteins extracted from cells or labeled for study of the interactions between biomolecules, are com- • Aromatic amino acids mon. Determination of protein concentration is a critical part of protein studies. In this application note, we use an Ocean HDX TECHNIQUES spectrometer to generate a standard curve for bovine serum • UV absorbance albumin (BSA). High sensitivity in the UV, ultra-low stray light performance and upgraded optics make the Ocean HDX ideal • Standard curve for UV absorbance measurements. generation Proteins are central to life. These intricate biomolecules play APPLICATIONS critical roles in cellular processes and metabolism, in catalyz- ing biochemical reactions and serving as structural elements, • Protein concentration and so much more. As vital elements for all life, proteins are determination often the subject of biomedical diagnostic studies and life sci- • Life sciences research ence research. Most research involving proteins, regardless of the answer being sought or the source of the protein, begins Unlocking the Unknown with Applied Spectral Knowledge. with quantifying the amount of protein present in • A(λ) is the absorption of the solution as a func- a sample. tion of wavelength • -
Beer-Lambert Law: Measuring Percent Transmittance of Solutions at Different Concentrations (Teacher’S Guide)
TM DataHub Beer-Lambert Law: Measuring Percent Transmittance of Solutions at Different Concentrations (Teacher’s Guide) © 2012 WARD’S Science. v.11/12 For technical assistance, All Rights Reserved call WARD’S at 1-800-962-2660 OVERVIEW Students will study the relationship between transmittance, absorbance, and concentration of one type of solution using the Beer-Lambert law. They will determine the concentration of an “unknown” sample using mathematical tools for graphical analysis. MATERIALS NEEDED Ward’s DataHub USB connector cable* Cuvette for the colorimeter Distilled water 6 - 250 mL beakers Instant coffee Paper towel Wash bottle Stir Rod Balance * – The USB connector cable is not needed if you are using a Bluetooth enabled device. NUMBER OF USES This demonstration can be performed repeatedly. © 2012 WARD’S Science. v.11/12 1 For technical assistance, All Rights Reserved Teacher’s Guide – Beer-Lambert Law call WARD’S at 1-800-962-2660 FRAMEWORK FOR K-12 SCIENCE EDUCATION © 2012 * The Dimension I practices listed below are called out as bold words throughout the activity. Asking questions (for science) and defining Use mathematics and computational thinking problems (for engineering) Constructing explanations (for science) and designing Developing and using models solutions (for engineering) Practices Planning and carrying out investigations Engaging in argument from evidence Dimension 1 Analyzing and interpreting data Obtaining, evaluating, and communicating information Science and Engineering Science and Engineering Patterns