Electrochemical Cells

Total Page:16

File Type:pdf, Size:1020Kb

Electrochemical Cells Analytical Electrochemistry: The Basic Concepts A. Electrochemical Cells 1. Defining the cell. The voltammetric methods we have described thus far involve the application of a potential from an external power source to a working electrode (WE) relative to that of a reference electrode (RE), and the measurement of current that flows as a result of that applied potential. In most instances, the electrochemical cell in which this process is carried out involves a solid working electrode immersed in a solution of charged electrolyte along with a reference half-cell of fixed composition. As was discussed in the Electrochemical thermodynamics section of this module, current passing in the cell as a result of applied 0.330 V potential can be of two basic types: 1) nonfaradaic, in which a Power supply change in the potential at the WE causes a rearrangement of electrolyte charges at each electrode/solution interface to counter the charge at the electrode surface; 2) faradaic, in which electrons are transferred across the electrode/solution interface to a solution species which is either oxidized or reduced. The two-electrode cell shown in Figure 32 is suitable as long as very little current passes during the experiment. This is the case for very low concentration solutions, or for very small working electrode dimensions (ultramicroelectrodes). In most practical applications, however, the circuit is designed to include a third electrode called the auxiliary electrode (AE). The WE RE device applying the potential and measuring the current (the potentiostat) electronically isolates the reference electrode so that little or no current passes through it, while allowing current to pass through Figure 32 the working and auxiliary electrodes. In this way, the activities of the reference half-cell components and thus the reference potential are prevented from changing during the experiment (a desirable characteristic!). The potentiostat will be discussed in a later section. 2. Positioning of the electrodes. For the remainder of our discussion, we will consider only the three-electrode electrochemical cell. The primary consideration for the relative location of the three electrodes in a cell is the minimization of the solution resistance. This is most usually accomplished by keeping the tips of the three electrodes as close together as feasible, while not interfering with the current paths between one another. Close approach is especially critical for the reference and working electrodes. The solution resistance between these electrodes leads to an iR drop that manifests itself as an error in the measured potential difference between them. In addition, the potentiostat is normally unable to electronically compensate for this resistance, as it can for the resistance between the reference electrode and the auxiliary electrode. A Luggin capillary arrangement is often employed, in which the RE is placed inside a tube drawn at the end to a fine capillary allowing very close positioning relative to the WE. A second important consideration involves the shape and size of the auxiliary electrode relative to the working electrode. The AE should be at least as large in area as the WE, and positioned symmetrically with respect to the WE so that the current density and potential experienced along its entire length is constant. In most voltammetric experiments, the three electrodes can be placed together in the same solution. Under certain conditions, however, either the AE or the RE (or both) may need to be physically isolated from the solution containing the WE. For example, if the goal of the experiment is the total electrochemical conversion of the bulk electroactive material to product (bulk electrolysis), products produced at the AE may be detrimental to the reaction occurring at the WE. Other times, small amounts of leakage of RE components, like Ag+ or even water from a Ag/AgCl reference, may cause undesirable reactions in the electrolyis solution, especially in experiments that involve non aqueous solvents. These problems can generally be avoided by the use of a glass frit or Vycor membrane (which allow charge to pass but not mixing of electrolyte) to separate the compartments of the electrochemical cell. 3. Other considerations. Most experiments require that dissolved oxygen be removed from the cell. Not only is it electrochemically active across the cathodic potential range, it is very likely to react with products formed by electron transfer. Typically, the solution is sparged with an inert gas like nitrogen or argon for 5 – 10 minutes prior to the experiment, and a “blanket” of inert gas maintained above the solution during the experiment. Cell volumes are quite variable and range from microliters to many mL, depending upon the goal of the experiment. Designs for cells can be quite simple – a glass vial with a screw cap drilled with holes for each electrode – to quite complex, requiring significant glass-blowing and/or machining skills. Cells are available from numerous electrochemical suppliers, some with very specific applications. Two common cell configurations for quiescent solution voltammetry are shown in Figure 33. On the left, the AE has been fashioned to be symmetrical to the WE. At right, a cell is shown with RE isolated from the solution containing the WE using a Luggin capillary. The AE is also isolated in a fritted compartment. Both cells have a fritted sparge tube to allow deoxygentation of the solution with inert gas. Figure 33 RE WE Sparge AE Sparge tube RE WE tube AE .
Recommended publications
  • 07 Chapter2.Pdf
    22 METHODOLOGY 2.1 INTRODUCTION TO ELECTROCHEMICAL TECHNIQUES Electrochemical techniques of analysis involve the measurement of voltage or current. Such methods are concerned with the interplay between solution/electrode interfaces. The methods involve the changes of current, potential and charge as a function of chemical reactions. One or more of the four parameters i.e. potential, current, charge and time can be measured in these techniques and by plotting the graphs of these different parameters in various ways, one can get the desired information. Sensitivity, short analysis time, wide range of temperature, simplicity, use of many solvents are some of the advantages of these methods over the others which makes them useful in kinetic and thermodynamic studies1-3. In general, three electrodes viz., working electrode, the reference electrode, and the counter or auxiliary electrode are used for the measurement in electrochemical techniques. Depending on the combinations of parameters and types of electrodes there are various electrochemical techniques. These include potentiometry, polarography, voltammetry, cyclic voltammetry, chronopotentiometry, linear sweep techniques, amperometry, pulsed techniques etc. These techniques are mainly classified into static and dynamic methods. Static methods are those in which no current passes through the electrode-solution interface and the concentration of analyte species remains constant as in potentiometry. In dynamic methods, a current flows across the electrode-solution interface and the concentration of species changes such as in voltammetry and coulometry4. 2.2 VOLTAMMETRY The field of voltammetry was developed from polarography, which was invented by the Czechoslovakian Chemist Jaroslav Heyrovsky in the early 1920s5. Voltammetry is an electrochemical technique of analysis which includes the measurement of current as a function of applied potential under the conditions that promote polarization of working electrode6.
    [Show full text]
  • Chapter 3: Experimental
    Chapter 3: Experimental CHAPTER 3: EXPERIMENTAL 3.1 Basic concepts of the experimental techniques In this part of the chapter, a short overview of some phrases and theoretical aspects of the experimental techniques used in this work are given. Cyclic voltammetry (CV) is the most common technique to obtain preliminary information about an electrochemical process. It is sensitive to the mechanism of deposition and therefore provides informations on structural transitions, as well as interactions between the surface and the adlayer. Chronoamperometry is very powerful method for the quantitative analysis of a nucleation process. The scanning tunneling microscopy (STM) is based on the exponential dependence of the tunneling current, flowing from one electrode onto another one, depending on the distance between electrodes. Combination of the STM with an electrochemical cell allows in-situ study of metal electrochemical phase formation. XPS is also a very powerfull technique to investigate the chemical states of adsorbates. Theoretical background of these techniques will be given in the following pages. At an electrode surface, two fundamental electrochemical processes can be distinguished: 3.1.1 Capacitive process Capacitive processes are caused by the (dis-)charge of the electrode surface as a result of a potential variation, or by an adsorption process. Capacitive current, also called "non-faradaic" or "double-layer" current, does not involve any chemical reactions (charge transfer), it only causes accumulation (or removal) of electrical charges on the electrode and in the electrolyte solution near the electrode. There is always some capacitive current flowing when the potential of an electrode is changing. In contrast to faradaic current, capacitive current can also flow at constant 28 Chapter 3: Experimental potential if the capacitance of the electrode is changing for some reason, e.g., change of electrode area, adsorption or temperature.
    [Show full text]
  • Square Wave Voltammetry
    Application Note S-7 Subject: Square Wave Voltammetry INTRODUCTION reverse pulse of the square wave occurs half way through the staircase step. The primary advantage of the pulse voltammetric techniques, such as normal pulse or differential pulse voltammetry, is their ability to discriminate against charging (capacitance) current. As a result, the pulse techniques are more sensitive to oxidation or reduction currents (faradaic currents) than conventional dc PLUS voltammetry. Differential pulse voltammetry yields peaks for faradaic currents rather than the sigmoidal waveform obtained with dc or normal pulse techniques. This results in improved resolution for multiple analyte systems and more convenient quantitation. Square wave voltammetry has received growing attention as a voltammetric technique for routine quantitative analyses. Although square wave voltammetry was E reported as long ago as 1957 by Barker1, the utility of the EQUALS technique was limited by electronic technology. Recent advances in both analog and digital electronics have made it feasible to incorporate square wave voltammetry into the PAR Model 394 Polarographic Analyzer. Square wave voltammetry can be used to perform an experiment much faster than normal and differential pulse techniques, which typically run at scan rates of 1 to 10 mV/sec. Square wave voltammetry employs scan rates TIME up to 1 V/sec or faster, allowing much faster determinations. A typical experiment requiring three minutes by normal or differential pulse techniques can be performed in a matter of seconds by square wave voltammetry. FIGURE 1: Applied excitation in square wave voltammetry. THEORY The timing and applied potential parameters for square The waveform used for square wave voltammetry is wave voltammetry are depicted in Figure 2.
    [Show full text]
  • Electrochemical Cells
    Electrochemical cells = electronic conductor If two different + surrounding electrolytes are used: electrolyte electrode compartment Galvanic cell: electrochemical cell in which electricity is produced as a result of a spontaneous reaction (e.g., batteries, fuel cells, electric fish!) Electrolytic cell: electrochemical cell in which a non-spontaneous reaction is driven by an external source of current Nils Walter: Chem 260 Reactions at electrodes: Half-reactions Redox reactions: Reactions in which electrons are transferred from one species to another +II -II 00+IV -II → E.g., CuS(s) + O2(g) Cu(s) + SO2(g) reduced oxidized Any redox reactions can be expressed as the difference between two reduction half-reactions in which e- are taken up Reduction of Cu2+: Cu2+(aq) + 2e- → Cu(s) Reduction of Zn2+: Zn2+(aq) + 2e- → Zn(s) Difference: Cu2+(aq) + Zn(s) → Cu(s) + Zn2+(aq) - + - → 2+ More complex: MnO4 (aq) + 8H + 5e Mn (aq) + 4H2O(l) Half-reactions are only a formal way of writing a redox reaction Nils Walter: Chem 260 Carrying the concept further Reduction of Cu2+: Cu2+(aq) + 2e- → Cu(s) In general: redox couple Ox/Red, half-reaction Ox + νe- → Red Any reaction can be expressed in redox half-reactions: + - → 2 H (aq) + 2e H2(g, pf) + - → 2 H (aq) + 2e H2(g, pi) → Expansion of gas: H2(g, pi) H2(g, pf) AgCl(s) + e- → Ag(s) + Cl-(aq) Ag+(aq) + e- → Ag(s) Dissolution of a sparingly soluble salt: AgCl(s) → Ag+(aq) + Cl-(aq) − 1 1 Reaction quotients: Q = a − ≈ [Cl ] Q = ≈ Cl + a + [Ag ] Ag Nils Walter: Chem 260 Reactions at electrodes Galvanic cell:
    [Show full text]
  • Supplementary Information
    Electronic Supplementary Material (ESI) for Journal of Materials Chemistry A This journal is © The Royal Society of Chemistry 2012 Supplementary Information 1. Synthesize the redox couples. The synthesis started commercially from available isothiocyanate which were transformed into the corresponding 1-ethyl-1H-tetrazole-5-thiol (ET) derivatives by cycloaddition reaction with sodium azide in refluxing ethanol according to the known procedures. Then the oxidized specie, bis(1-ethyltetrazol-5-yl) (BET) disulfides, was prepared by oxidation of the corresponding ET with hydrogen peroxide, and thiolate (ET-) form was obtained by deprotonation of the corresponding mercaptan (ET) with sodium bicarbonate. The structures of the above-mentioned redox couple was proved by the combination of 1HNMR spectroscopy, mass spectroscopy (ESI-MS) and elemental analyses. The NMR spectra were recorded at 298 K in CDCl3 at 300 MHz on a Varian Mercury-VX300 spectrometer. The chemical shifts were recorded in parts per million (ppm) with TMS as the internal reference. ESI mass spectra were determined using Finnigan LCQ Advantage mass spectrometer. Elemental analyses were performed with Thermo Quest Flash EA1112. The electrolyte consisted of 0.4 M of ET-, 0.05 M of BET, 0.4 M 18-crown-6 (18-C-6), 0.05 M LiClO4 and 0.5 M 4-tert-butylpyridine (TBP) in acetonitrile (ACN). 2. Preparation of electrodes. The NiS electrodes were electrodeposited onto a fluorine-doped tin oxide (FTO) glass substrate (13Ω/□) from an aqueous electrolyte consisting of 1M CH4N2S and 40mM NiCl2∙6H2O in a single-compartment glass cell with three-electrode configuration using electrochemical work station.
    [Show full text]
  • 3 PRACTICAL APPLICATION BATTERIES and ELECTROLYSIS Dr
    ELECTROCHEMISTRY – 3 PRACTICAL APPLICATION BATTERIES AND ELECTROLYSIS Dr. Sapna Gupta ELECTROCHEMICAL CELLS An electrochemical cell is a system consisting of electrodes that dip into an electrolyte and in which a chemical reaction either uses or generates an electric current. A voltaic or galvanic cell is an electrochemical cell in which a spontaneous reaction generates an electric current. An electrolytic cell is an electrochemical cell in which an electric current drives an otherwise nonspontaneous reaction. Dr. Sapna Gupta/Electrochemistry - Applications 2 GALVANIC CELLS • Galvanic cell - the experimental apparatus for generating electricity through the use of a spontaneous reaction • Electrodes • Anode (oxidation) • Cathode (reduction) • Half-cell - combination of container, electrode and solution • Salt bridge - conducting medium through which the cations and anions can move from one half-cell to the other. • Ion migration • Cations – migrate toward the cathode • Anions – migrate toward the anode • Cell potential (Ecell) – difference in electrical potential between the anode and cathode • Concentration dependent • Temperature dependent • Determined by nature of reactants Dr. Sapna Gupta/Electrochemistry - Applications 3 BATTERIES • A battery is a galvanic cell, or a series of cells connected that can be used to deliver a self-contained source of direct electric current. • Dry Cells and Alkaline Batteries • no fluid components • Zn container in contact with MnO2 and an electrolyte Dr. Sapna Gupta/Electrochemistry - Applications 4 ALKALINE CELL • Common watch batteries − − Anode: Zn(s) + 2OH (aq) Zn(OH)2(s) + 2e − − Cathode: 2MnO2(s) + H2O(l) + 2e Mn2O3(s) + 2OH (aq) This cell performs better under current drain and in cold weather. It isn’t truly “dry” but rather uses an aqueous paste.
    [Show full text]
  • Hydrodynamic Electrodes and Microelectrodes
    CHEM465/865, 2004-3, Lecture 20, 27 th Sep., 2004 Hydrodynamic Electrodes and Microelectrodes So far we have been considering processes at planar electrodes. We have focused on the interplay of diffusion and kinetics (i.e. charge transfer as described for instance by the different formulations of the Butler-Volmer equation). In most cases, diffusion is the most significant transport limitation. Diffusion limitations arise inevitably, since any reaction consumes reactant molecules. This consumption depletes reactant (the so-called electroactive species) in the vicinity of the electrode, which leads to a non-uniform distribution (see the previous notes). ______________________________________________________________________ Note: In principle, we would have to consider the accumulation of product species in the vicinity of the electrode as well. This would not change the basic phenomenology, i.e. the interplay between kinetics and transport would remain the same. But it would make the mathematical formalism considerably more complicated. In order to simplify things, we, thus, focus entirely on the reactant distribution, as the species being consumed. ______________________________________________________________________ In this part, we are considering a semiinfinite system: The planar electrode is assumed to have a huge surface area and the solution is considered to be an infinite reservoir of reactant. This simple system has only one characteristic length scale: the thickness of the diffusion layer (or mean free path) δδδ. Sometimes the diffusion layer is referred to as the “Nernst layer” . Now: let’s consider again the interplay of kinetics and diffusion limitations. Kinetic limitations are represented by the rate constant k 0 (or equivalently by the 0=== 0bα b 1 −−− α exchange current density j nFkcred c ox ).
    [Show full text]
  • Elements of Electrochemistry
    Page 1 of 8 Chem 201 Winter 2006 ELEM ENTS OF ELEC TROCHEMIS TRY I. Introduction A. A number of analytical techniques are based upon oxidation-reduction reactions. B. Examples of these techniques would include: 1. Determinations of Keq and oxidation-reduction midpoint potentials. 2. Determination of analytes by oxidation-reductions titrations. 3. Ion-specific electrodes (e.g., pH electrodes, etc.) 4. Gas-sensing probes. 5. Electrogravimetric analysis: oxidizing or reducing analytes to a known product and weighing the amount produced 6. Coulometric analysis: measuring the quantity of electrons required to reduce/oxidize an analyte II. Terminology A. Reduction: the gaining of electrons B. Oxidation: the loss of electrons C. Reducing agent (reductant): species that donates electrons to reduce another reagent. (The reducing agent get oxidized.) D. Oxidizing agent (oxidant): species that accepts electrons to oxidize another species. (The oxidizing agent gets reduced.) E. Oxidation-reduction reaction (redox reaction): a reaction in which electrons are transferred from one reactant to another. 1. For example, the reduction of cerium(IV) by iron(II): Ce4+ + Fe2+ ! Ce3+ + Fe3+ a. The reduction half-reaction is given by: Ce4+ + e- ! Ce3+ b. The oxidation half-reaction is given by: Fe2+ ! e- + Fe3+ 2. The half-reactions are the overall reaction broken down into oxidation and reduction steps. 3. Half-reactions cannot occur independently, but are used conceptually to simplify understanding and balancing the equations. III. Rules for Balancing Oxidation-Reduction Reactions A. Write out half-reaction "skeletons." Page 2 of 8 Chem 201 Winter 2006 + - B. Balance the half-reactions by adding H , OH or H2O as needed, maintaining electrical neutrality.
    [Show full text]
  • Μstat 4000P Multi Potentiostat
    µStat 4000P Multi Potentiostat 01 Ref. STAT4000P Following the format of our multipotentiostats with a size of only 22x20x7 cm, includes 4 channels that can act at the same time as 4 independent potentiostats; it also includes one multichannel that can act as a poten- tiostat where up to 4 working electrodes share an auxiliary and a reference electrode. With µStat 4000P users can perform up to 4 different electrochemical techniques at the same time; or carry out the study of one technique’s parameter in just one step by applying the same electrochemical technique in several channels but selecting different values for the parameter under study. These are just exam- ples of the enormous capabilities that our new instrument offers. µStat 4000P can be applied for Voltammetric or Amperometric measurements, including 12 electroanalytical techniques. In addition, µStat 4000P owners can later upgrade their instrument to a µStat 4000P by just purchasing an extension. This self-upgrade does not require any hardware modification, but it is implemented by means of a Galvanostat software update kit. This Multi Potentiostat is Li-ion Battery powered (DC charger adaptor also compatible), and can be easily connected to a PC via USB or through Wireless connection. µStat 4000P is controlled by the powerful software “DropView 8400” which is included and that allows plotting of the measurements and performing the analysis of results. DropView software provides powerful functions such as experimental control, graphs or file handling, among others. Available
    [Show full text]
  • Physical and Analytical Electrochemistry: the Fundamental
    Electrochemical Systems The simplest and traditional electrochemical process occurring at the boundary between an electronically conducting phase (the electrode) and an ionically conducting phase (the electrolyte solution), is the heterogeneous electro-transfer step between the electrode and the electroactive species of interest present in the solution. An example is the plating of nickel. Ni2+ + 2e- → Ni Physical and Analytical The interface is where the action occurs but connected to that central event are various processes that can Electrochemistry: occur in parallel or series. Figure 2 demonstrates a more complex interface; it represents a molecular scale snapshot The Fundamental Core of an interface that exists in a fuel cell with a solid polymer (capable of conducting H+) as an electrolyte. of Electrochemistry However, there is more to an electrochemical system than a single by Tom Zawodzinski, Shelley Minteer, interface. An entire circuit must be made and Gessie Brisard for measurable current to flow. This circuit consists of the electrochemical cell plus external wiring and circuitry The common event for all electrochemical processes is that (power sources, measuring devices, of electron transfer between chemical species; or between an etc). The cell consists of (at least) electrode and a chemical species situated in the vicinity of the two electrodes separated by (at least) one electrolyte solution. Figure 3 is electrode, usually a pure metal or an alloy. The location where a schematic of a simple circuit. Each the electron transfer reactions take place is of fundamental electrode has an interface with a importance in electrochemistry because it regulates the solution. Electrons flow in the external behavior of most electrochemical systems.
    [Show full text]
  • Redox Potential Measurements M.J
    Redox Potential Measurements M.J. Vepraskas NC State University December 2002 Redox potential is an electrical measurement that shows the tendency of a soil solution to transfer electrons to or from a reference electrode. From this measurement we can estimate whether the soil is aerobic, anaerobic, and whether chemical compounds such as Fe oxides or nitrate have been chemically reduced or are present in their oxidized forms. Making these measurements requires three basic pieces of equipment: 1. Platinum electrode 2. Voltmeter 3. Reference electrode The basic set-up is shown below: Voltmeter 456 mv Soil surface Reference Pt wire electrode Fig. 1. The Pt wire is buried into the soil to be in contact with the soil solution. The reference electrode must also be in contact with the soil solution. It has a ceramic tip, which can be placed in the soil, or the electrode can be placed in a salt bridge, which is itself placed in the soil. Wires from both the Pt electrode and reference electrode are connected to the voltmeter. Pt Electrodes Platinum electrodes consist of a small piece of platinum wire that is soldered or fused to wire made another metal. Platinum conducts electrons from the soil solution to the wire to which it is attached. Platinum is used because it is assumed to be an inert metal. This means it does not give up its own electrons (does not oxidize) to the wire or soil solution. Iron containing materials such as steel will oxidize themselves and send their own electrons to the voltmeter. As a result the voltage we measure will not result solely from electrons being transferred to or from the soil.
    [Show full text]
  • Advances in Materials Design for All-Solid-State Batteries: from Bulk to Thin Films
    applied sciences Review Advances in Materials Design for All-Solid-state Batteries: From Bulk to Thin Films Gene Yang 1, Corey Abraham 2, Yuxi Ma 1, Myoungseok Lee 1, Evan Helfrick 1, Dahyun Oh 2,* and Dongkyu Lee 1,* 1 Department of Mechanical Engineering, College of Engineering and Computing, University of South Carolina, Columbia, SC 29208, USA; [email protected] (G.Y.); [email protected] (Y.M.); [email protected] (M.L.); [email protected] (E.H.) 2 Chemical and Materials Engineering Department, Charles W. Davidson College of Engineering, San José State University, San José, CA 95192-0080, USA; [email protected] * Correspondence: [email protected] (D.O.); [email protected] (D.L.) Received: 15 June 2020; Accepted: 7 July 2020; Published: 9 July 2020 Featured Application: All solid-state lithium batteries, all solid-state thin-film lithium batteries. Abstract: All-solid-state batteries (SSBs) are one of the most fascinating next-generation energy storage systems that can provide improved energy density and safety for a wide range of applications from portable electronics to electric vehicles. The development of SSBs was accelerated by the discovery of new materials and the design of nanostructures. In particular, advances in the growth of thin-film battery materials facilitated the development of all solid-state thin-film batteries (SSTFBs)—expanding their applications to microelectronics such as flexible devices and implantable medical devices. However, critical challenges still remain, such as low ionic conductivity of solid electrolytes, interfacial instability and difficulty in controlling thin-film growth. In this review, we discuss the evolution of electrode and electrolyte materials for lithium-based batteries and their adoption in SSBs and SSTFBs.
    [Show full text]