news and views

Microfluidics closes in on point-of-care assays

Peter K Sorger

An integrated microfluidic device enables rapid detection of protein biomarkers in patient samples.

Routine assessment of biomarkers promises to biomarkers from the laboratory to the clinic small feature size (such as resonance-based transform many areas of medicine, from clini- has been remarkably difficult, in part because mass sensing7) and to integrate multiple cal trials, which will benefit from - relatively large trials are required to assess their separators and sensors8. Integration is widely based segmentation of patient populations, utility and because techniques that are routine thought to be of particular importance, and the to the diagnosis and treatment of complex in biomarker research, such as mass spectrom- integrated blood barcode chip from Fan et al.3 disease1. To add real value, biomarkers must etry, are not common in clinical laboratories nicely demonstrates this capability (Fig. 1). be measurable by clinicians and provide use- (see ref. 5 for a discussion of the technical and The chip first separates plasma from blood 2 3 http://www.nature.com/naturebiotechnology ful new information . In this issue, Fan et al. commercial challenges). Such translational using microchannels and then assays for tackle the first of these challenges by develop- bottlenecks should be eased by the availabil- specific plasma components in the same chan- ing a microfluidic integrated blood barcode ity of simple and rugged biomarker nels. Isolation of plasma is achieved based on chip that could serve as a point-of-care device devices6. the tendency of cells to enter the higher-flow for measuring multiple protein biomarkers in Several features of microfluidic devices sug- microchannel when a bifurcation is encoun- small samples of blood. gest that they would be ideal for point-of-care tered (as first observed in rat capillaries9). It is increasingly clear that the success of assays. These include their ability to process On-chip separation of plasma from blood, pre- future therapeutics will depend on identifying very small samples using small quantities of viously described by Yang et al.10, obviates the those patients who are most likely to respond. reagents (perhaps stored on-chip), to exploit need for prior centrifugation of the sample. In Both payers and patients will demand evidence the physics of low-Reynolds-number environ- the integrated blood barcode chip, the plasma- that drugs are working in specific individuals. ments (such as free-flow electrophoresis) and skimming channels contain successive rows of Sophisticated biomarker analysis could con- tribute markedly to achieving these goals, Nature Publishing Group Group 200 8 Nature Publishing CCellsells

© improving the diagnosis, therapy and man- agement of . Indeed, for good or ill, companies such as Biophysical in Austin, Texas, On-chip are already marketing panels of diagnostic bio- storage BIOMARKE markers directly to patients. LEVEL R Biomarkers include proteins, hormones or other biomolecules whose levels, states or activ- ities change with disease state or treatment and provide diagnostic or prognostic information. Ideally, such biomarkers would be mechanisti- Plasma cally linked to the etiology or pathobiology of Blood the disease, but most biomarkers simply show ~5 min ~60 min ~15 min tissue-specific patterns of expression or statisti- Ris-Vicari Katie cal correlation with a disease state. For exam- ple, prostate-specific antigen is a tissue-specific Figure 1 Possible workflow for a commercial microfluidic biomarker assay system. In devices such as the integrated blood barcode chip3, chip regeneration and readout could involve an external workstation glycoprotein found on both transformed and that includes pumps and wash buffers, although some chip-specific reagents might be stored on board. normal prostate epithelia that has proven use- A blood sample collected with a standard lancet is introduced into the workstation and diluted with ful in monitoring cancer recurrence and, more EDTA to prevent clotting. The blood is then introduced into the microfluidic chip, where it is separated controversially, in diagnosis4. However, moving into cells and plasma; in the integrated blood barcode chip cells go to waste, but they could be recovered for detection of genetic biomarkers and other purposes. Multiple analytes are then captured from plasma Peter K. Sorger is in the Department of Systems by patterned immobilized through DNA hybridization; alternatively, or other affinity capture reagents might be used. After washing and introduction of secondary reagents, analyte levels Biology, WAB Room 438, Harvard Medical are read using the workstation. Biomarker measurements must be carefully integrated with patient School, 200 Longwood Avenue, Boston, history using sophisticated statistical algorithms because biomarker data are rarely simple or univariant. Massachusetts 02115, USA. Ultimately, it would be desirable to produce devices in which more of the instrumentation was on board e-mail: [email protected] the chip so that a handheld device could be used to prepare and read the chip.

nature volume 26 number 12 DECEMBER 2008 1345 news and views

immobilized antibodies that capture analytes much higher nonspecific binding than that of whose reliability, simplicity and low cost will present in the plasma for subsequent detection serum13. Further experiments with fresh blood revolutionize the use of protein biomarkers. using fluorescent secondary antibodies. from healthy volunteers and patients, as out- 3 1. Braunwald, E. N. Engl. J. Med. 358, 2148–2159 Fan et al. show that the levels of prostate- lined by the authors, should resolve whether (2008). specific antigen detected in patient by the rapid processing of blood made possible 2. Morrow, D.A. & de Lemos, J.A. Circulation 115, 949– their chip correlate well with those measured by their chip will solve the sensitivity problems 952 (2007). 3. Fan, R. et al. Nat. Biotechnol. 26, 1373–1378 by conventional techniques, and the overall that have impaired less sophisticated devices. (2008). sensitivity of the device is sufficient to quan- It is now clear that the micro total analysis 4. Barry, M.J. N. Engl. J. Med. 344, 1373–1377 (2001). 5. Kling, J. Nat. Biotechnol. 24, 891–893 (2006). tify multiple proteins and cytokines at levels system (µTAS or lab-on-a-chip) proposed over 6. Goluch, E.D. et al. Lab Chip 6, 1293–1299 (2006). 14 relevant to human disease. However, whether a decade ago will be substantially harder to 7. Burg, T.P. et al. Nature 446, 1066–1069 (2007). fluorescence will ultimately be the best way achieve than was first envisioned. Indeed, sev- 8. El-Ali, J., Sorger, P.K. & Jensen, K.F. Nature 442, 403– 411 (2006). to detect biomarkers remains unclear; a wide eral companies founded on the µTAS concept 9. Svanes, K. & Zweifach, B.W. Microvasc. Res. 1, 210– variety of detection methods are under devel- have already faltered or failed. Looking forward, 220 (1968). opment, including several that are label-free7. we can nonetheless expect steady improve- 10. Yang, S., Undar, A. & Zahn, J.D. Lab Chip 6, 871–880 (2006). Nearly a decade after widespread clinical use ment in microfluidic technologies, antibody- 11. Boozer, C., Ladd, J., Chen, S. & Jiang, S. Anal. Chem. of was first anticipated, the field and -based capture of biomolecules, 78, 1515–1519 (2006). 12. Bailey, R.C. et al. J. Am. Chem. Soc. 129, 1959–1967 remains plagued by technical problems. One label-free detection and on-chip integration (2007). issue concerns the sensitivity and durability of of multiple preparative and analytical units 13. Vaisocherova, H. et al. Anal. Chem. 80, 7894–7901 antibody arrays, on which most biochemical with diverse functions. It seems highly likely (2008). 14. van den Berg, A. & Lammerink, T.S.J. in Microsystem assays rely. The integrated blood barcode chip that devices such as the integrated blood bar- Technology in Chemistry and Life Sciences (eds. Manz, addresses this problem by using DNA-directed code chip presage commercial analytic devices A. & Becker, H.) pp. 21–50 (Springer, Berlin, 1999). protein immobilization11, in which a surface coated with single-stranded DNA is hybridized

http://www.nature.com/naturebiotechnology to antibodies chemically conjugated to comple- mentary DNA. This seemingly baroque scheme has important advantages over direct antibody fine-tune oncolytic conjugation, including reliance on robust and stable DNA-glass surfaces during device fab- viruses rication and storage, formation of delicate antibody-containing features just before use, John C Bell & David Kirn and creation of capture features with densely packed antigen-combining domains, thereby Targeting by tissue-specific microRNAs enhances the efficacy and safety of 12 increasing sensitivity . tumor-killing viruses. Passivation and the prevention of fouling are two additional challenges in using microfluidic devices with biological samples. Passivation Gene silencing by endogenous microRNAs transcriptional regulation through specific Nature Publishing Group Group 200 8 Nature Publishing

© refers to the modification of surfaces so as to (miRNAs) has recently been exploited to recognition of short sequences, often located minimize nonspecific binding, and fouling to control the tropism of gene-therapy vec- in the 3′ untranslated region (UTR), in target unwanted binding to device surfaces and to tors. By including the target sequence of a mRNAs6. Depending on its degree of com- plugging of microchannels by components tissue-specific miRNA in the genome of len- plementarity to the target, the miRNA can of the sample. Fouling leads to irreproduc- tiviral vectors, Naldini and colleagues sup- affect either the stability or translation of the ible measurements and device failure. Because pressed transgene expression specifically mRNA7. miRNAs have complex expression most microfluidic devices based on soft lithog- in selected cell types such as hematopoietic profiles that reflect the important roles they raphy (such as the integrated blood barcode cells and hepatocytes1,2. This strategy has play in the control of mammalian growth chip) are bonded to glass, and glass has a very now been extended to control targeting of and development. The human genome is esti- high nonspecific binding capacity for proteins, oncolytic viruses. The new studies, appearing mated to contain >500 miRNA genes. Some and other biomolecules, effective pas- in Molecular Therapy3, Journal of Virology4 are expressed in a tissue-specific fashion, sivation is difficult to achieve. Perfect device and Nature Medicine5, mark the beginning whereas others are constitutively expressed passivation and prefiltering of samples, which of innovative efforts aimed at discovering or turned on in response to endogenous would solve the problem of fouling, has never combinations of viruses and miRNA targets cues or stress signals8. Another important been achieved. Fortunately, the multilayer that yield safer and more effective anticancer function of miRNAs, clearly demonstrated polyamine-DNA-antibody surfaces created by virotherapeutics. in plants and invertebrates9, is to suppress DNA-encoded antibody immobilization are miRNAs are versatile, noncoding RNAs, viruses by binding to cognate sequences in anti-fouling and appear to be very effectively ~22 nucleotides in length, that exert post- viral mRNAs. passivated. The remaining fouling problems Oncolytic viruses are vectors engineered or relate to background that limits the ultimate John C. Bell is in the Ottawa Health Research selected to infect and kill cancer cells while sensitivity of the device. Institute, Ottawa, Ontario K1G 0K8, Canada leaving normal cells relatively unscathed10,11. Fan et al.3 demonstrate impressive sensitivity and David Kirn is at Jennerex Inc., One Market Clinical data show that this class of therapeu- with their chip, but largely using serum (plasma St., Spear Tower, Suite 2260, San Francisco, tic is safe and cancer selective but also suggest cleared of fibrinogen and other clotting fac- California 94105, USA. that more potent viruses—targeted specifi- tors) rather than plasma itself, which gives e-mail: [email protected] or [email protected] cally to malignancies—would be desirable to

1346 volume 26 number 12 DECEMBER 2008 nature biotechnology