Policy # QEQMI03001 Page 1 of 12 Department of Microbiology Quality Manual Version: 1.5 CURRENT Section: Equipment Subject Title: Equipment Maintenance Procedure Prepared by QA Committee Issued by: Manager Revision Date: 5/28/2021 Approved by Laboratory Director: Review Date: 5/28/2023 Microbiologist-in-Chief Uncontrolled When Printed

Purpose:

To provide instructions for the general maintenance of all equipment in the department.

Responsibility:

All Microbiology staff is responsible for maintaining good working condition of all laboratory equipment.

Procedure:

A. Routine Maintenance Performed by Laboratory Staff Laboratory staff perform maintenance as directed in the Standard Operating Procedure (SOP) specific for each instrument. The completion of maintenance is documented on the analyzer history logs and / or micQC.

B. Preventive Maintenance by Company Representative The frequency at which PM is completed is determined by the laboratory contract with the associated company. PM is performed by contracted companies.

Senior technologists responsible for arranging and ensuring that company representatives visit the laboratory to perform PMs before they are due. To support the senior technologists an automated alert system has been set. A member from the Quality Improvement (QI) committee and charge technologist will be sent an alert 1 month before each scheduled PM for each piece of equipment. The QI committee member will notify the senior technologists of the upcoming PM due and follow up with the alerts until PM is completed. The Charge technologist receiving the same alerts is a back up to the QI member and will monitor the process.

Once a PM is completed, a sticker indicating the next PM date is applied on the equipment. A record of the PM is provided by the company to the QI committee member and is saved to the T drive. Records of PM are kept as per the record retention time document.

C. Unscheduled Service by Company Representative In the case of service needed outside scheduled provision by the company, senior technologists or bench technologists will label equipment as “Out of service” or “not in use” as appropriate, then contact the company by the number provided in the chart below.

UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY

NOTE: This document is Uncontrolled When Printed. Any documents appearing in paper form that are not stamped in red "MASTER COPY" are not controlled and should be checked against the document (titled as above) on the server prior to use. Management System\UHN_Mount Sinai Hospital Microbiology\Quality Manual Policies and Procedures\Equipment\

Policy # QEQMI03001 Page 2 of 12 Department of Microbiology Quality Manual Version: 1.5 CURRENT Section: Equipment Subject Title: Equipment Maintenance Procedure

Equipment Maintenance Item Frequency Maintenance by Accumet AB150 pH check When used Technologist AltoStar As per Analyzer history log Daily Technologist QEQMI03001hAltostar As per Analyzer history log Weekly Technologist QEQMI03001hAltostar Unscheduled Service As Required Altona/Hamilton Preventive Maintenance Support 1-800-527-5269 Use ATTEST When used Technician MSH: Save printouts Unscheduled Service MSH As Required McDavis Sales and Preventive Maintenance Quarterly Services Tel: 416-493-6601 Toll: 1-800-274-2122 Architect See Architect Maintenance Daily, Weekly, Bench Technologist Log Monthly Unscheduled Service As Required Abbott Preventive Maintenance Annual Customer Support Tel: 905-858-2450 Luminex Aries Discard used cassettes Daily Technologist Disinfect sample trays End of day cleaning: Magazines, Barcode Reader, Keyboad, Touchscreen Insert empty magazine Weekly Technologist Reboot system Unscheduled Downtime As Requred Luminex Corporation Preventive Maintenance Semi-Annual Tel: 1-877-785-2323 Balance Maintenance Annual IDEAL Unscheduled Service As Required Unity Lab Services Preventive Maintenance Annual Toll: 1-800-234-7437 BioFire Area and Instrument Daily Technologist cleaning/disinfecting Waste discard Unscheduled Maintenance As Required Technical Support Preventive Maintenance N/A

UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY

NOTE: This document is Uncontrolled When Printed. Any documents appearing in paper form that are not stamped in red "MASTER COPY" are not controlled and should be checked against the document (titled as above) on the server prior to use. Management System\UHN_Mount Sinai Hospital Microbiology\Quality Manual Policies and Procedures\Equipment\

Policy # QEQMI03001 Page 3 of 12 Department of Microbiology Quality Manual Version: 1.5 CURRENT Section: Equipment Subject Title: Equipment Maintenance Procedure

Equipment Maintenance Item Frequency Maintenance by Bio-Rad Clean external surfaces Daily Technologist Clean work area Dispose of waste Run CFX Qualificate Plate Annual Technologist Unscheduled Maintenance As Required BioRad Tel: 905-364-3435 Toll: 1-800-268-0213 Biosafety Cabinets Disinfect work surface & Daily First user of the day Check air pressure Disinfect entire hood Bi-Monthly First user the 1st and 15th day of the month Decontamination, Testing, Annual Certification Raeyco Unscheduled Service As Required Tel: 877-772-3926 Preventive Maintenance Annual Block Heater/Dry Bath Measure Temperature Daily Bench Technologist

Preventive Maintenance Annual Unity Lab Services Toll: 1-800-234-7437 Centrifuge Clean Daily Bench Assigned RPM Check Quarterly Charge Technologist Inspection including Annually Fisher Scientific Calibration And Certification Preventive Maintenance Annual Unity Lab Services Toll: 1-800-234-7437 Chef II/III Change Buffer Weekly Technologist Change Probes As Required Technologist

UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY

NOTE: This document is Uncontrolled When Printed. Any documents appearing in paper form that are not stamped in red "MASTER COPY" are not controlled and should be checked against the document (titled as above) on the server prior to use. Management System\UHN_Mount Sinai Hospital Microbiology\Quality Manual Policies and Procedures\Equipment\

Policy # QEQMI03001 Page 4 of 12 Department of Microbiology Quality Manual Version: 1.5 CURRENT Section: Equipment Subject Title: Equipment Maintenance Procedure

Equipment Maintenance Item Frequency Maintenance by COBAS system  Clean initialization post Daily Technologist  Clean D-cup handler tip  Check and fill wash buffer reservoir  Check and empty waste container  Prime the analyzer  Check syringes, tubing, and valve block  Check transfer tip  Check printer  UV decontamination  Perform extended prime Weekly Technologist  Clean trays/racks Monthly/As Technologist  Replace filters Required  Replace syringe tips and O- rings  Replace transfer tip  Replace Taqman lamp Unscheduled Service As Required Roche Preventive Maintenance Annual Tel: 1-877-273-3433 COBAS 6800  Check, Load / Unload Daily Technologist consumables  Clean reagent and Weekly Technologist consumable drawers  Clean sample supply module  Periodic maintenance  Wipe internal plastic cover compartments  Decontaminate racks/trays Unscheduled Service As Required Roche Preventive Maintenance Bi-Annual Tel: 1-877-273-3433 EasyMag  Inspect drip tray, reagent Daily Technologist filters, carbon filter  Empty/rinse waste  Clean surfaces UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY

NOTE: This document is Uncontrolled When Printed. Any documents appearing in paper form that are not stamped in red "MASTER COPY" are not controlled and should be checked against the document (titled as above) on the server prior to use. Management System\UHN_Mount Sinai Hospital Microbiology\Quality Manual Policies and Procedures\Equipment\

Policy # QEQMI03001 Page 5 of 12 Department of Microbiology Quality Manual Version: 1.5 CURRENT Section: Equipment Subject Title: Equipment Maintenance Procedure

Equipment Maintenance Item Frequency Maintenance by  Clean dispense probe Weekly Technologist program  Clean dispense probe clean Bi-Annual Biomerieux station, aspirator Support Service disposable o-ring, magnet Toll: 1-800-361-7321 array, reagent caps and [email protected] connectors, sample vessel, carrier box.  Preventive Maintenance Bi-Annual Epimotion  Clean external surfaces Daily Technologist including screen/mouse/keypad  Decontaminate sample and reagent racks  Discard used tips and Weekly Technologist decontaminate tip holder  Preventive Maintenance Annually Eppendorf Canada Toll: 1-800-263-8715 Evolis  Check system liquid level Daily Technologist  Inspect for stains and spills  Wide the tip adapter  Empty liquid and tip waste  Perform Washer Clean Weekly Technologist  Decontaminate pipettor wash station, tip ejection slide  Clean instrument surfaces  Run Washer Manifold Monthly Technologist Disinfect assay  Decontaminate sample and reagent racks, plate carrier, system liquid container.  Clean wash buffer . Unscheduled Service As Required Biorad Laboratories Preventive Maintenance Annually Toll: 1-800-361-1808 Freezer Measure Temperature Daily Phillips Monitoring System

UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY

NOTE: This document is Uncontrolled When Printed. Any documents appearing in paper form that are not stamped in red "MASTER COPY" are not controlled and should be checked against the document (titled as above) on the server prior to use. Management System\UHN_Mount Sinai Hospital Microbiology\Quality Manual Policies and Procedures\Equipment\

Policy # QEQMI03001 Page 6 of 12 Department of Microbiology Quality Manual Version: 1.5 CURRENT Section: Equipment Subject Title: Equipment Maintenance Procedure

Equipment Maintenance Item Frequency Maintenance by Clean and Defrost if Required Bi-Annual Technician Unscheduled Service As Required Gibson Air Preventive Maintenance Quarterly Tel: 416-447-1800 Fume Hoods Disinfect work surface & Daily First user of the day Check air pressure Daily Maintenance & Disinfect Bi-Monthly First user the 1st and entire hood 15th day of the month Decontamination, Testing, Bi-Annual Raeyco Certification Tel: 877-772-3926 Unscheduled Service As Required Gel Doc XR+ Clean surface Weekly Technologist

GeneXpert  Disinfect instrument Daily Technician surfaces and work bench  Disinfect the instrument Monthly Technician surfaces, cartridge bay interior, plunger rod  Clean fan filters Recalibration Annually Intermedico Unscheduled Service As Required Tell: 905-470-2520 Preventive Maintenance Annual Toll: 1-800-387-9643 [email protected] Glass washer Unscheduled Service As Required McDavis Sales and Preventive Maintenance Quarterly Services Tel: 416-493-6601 Toll: 1-800-274-2122 Hot air oven Unscheduled Service As Required McDavis Sales and Preventive Maintenance Quarterly Services Tel: 416-493-6601 Toll: 1-800-274-2122 – CO2 Measure Temperature Daily Phillips Monitoring System CO2 level recording Daily Technician  CO2 level measuring Weekly Technician (Fryite)

UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY

NOTE: This document is Uncontrolled When Printed. Any documents appearing in paper form that are not stamped in red "MASTER COPY" are not controlled and should be checked against the document (titled as above) on the server prior to use. Management System\UHN_Mount Sinai Hospital Microbiology\Quality Manual Policies and Procedures\Equipment\

Policy # QEQMI03001 Page 7 of 12 Department of Microbiology Quality Manual Version: 1.5 CURRENT Section: Equipment Subject Title: Equipment Maintenance Procedure

Equipment Maintenance Item Frequency Maintenance by Clean and disinfect Biweekly Technician  Incubator Minimum  Shelves  Humidity pan Replace distilled water in humidity pan

Unscheduled Service As Required Unity Lab Services Preventive Maintenance Annual Tel: 1-800-234-7437 Incubator – O2 Measure Temperature Daily Phillips Monitoring System Clean and disinfect Biweekly Technician  Incubator Minimum  Shelves  Humidity pan Replace distilled water in humidity pan Unscheduled Service As Required Unity Lab Services Preventive Maintenance Annual Tel: 1-800-234-7437 Isoplator  Clean aluminum plate Daily Specimen guides Management  Lightly sandpaper buildup on loops  Swab clean any buildup under loop holders Unscheduled As Required Unity Lab Services Preventive Maintenance Semi-Annual Tel: 1-800-234-7437 Light Cycler Perform self test Daily (once per Technologist 24 hour period) Luminex NxTag Check fluid levels Daily Technologist Enhanced start-up Verification Clean fluid reservoir Routine system shutdown Clean probe Weekly Technologist Visual inspection for leaks Set probe height Enhanced start-up UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY

NOTE: This document is Uncontrolled When Printed. Any documents appearing in paper form that are not stamped in red "MASTER COPY" are not controlled and should be checked against the document (titled as above) on the server prior to use. Management System\UHN_Mount Sinai Hospital Microbiology\Quality Manual Policies and Procedures\Equipment\

Policy # QEQMI03001 Page 8 of 12 Department of Microbiology Quality Manual Version: 1.5 CURRENT Section: Equipment Subject Title: Equipment Maintenance Procedure

Equipment Maintenance Item Frequency Maintenance by Calibration/Verification Wipe exterior surfaced Unscheduled Downtime As Required Luminex Corporation Preventive Maintenance Annual Tel: 1-877-785-2323 Max Q4000 Clean surface Weekly Technologist Preventive maintenance Annual Unity Lab Services Toll: 1-800-234-7437 [email protected] Kohler Illumination Daily Technologist Complete cleaning, Semi-Annual lubrication, Calibration and Raeyco minor repairs Tel: 604-444-0004 Unscheduled Service As Required Preventive Maintenance Bi-Annual Inspect and Calibrate Semi-Annual Techno Laboratory Unscheduled Service As Required Consulting Inc. Tel: 905-308-1535 [email protected] Refrigerators Measure Temperature Daily Phillips Monitoring System Seegene Starlet Unscheduled Service As Required Jaesun Kim (Primary Engineer) Preventive Maintenance Bi-Annual Jaesunkim0503@seeg ene.com

John Choi (Secondary Engineer) [email protected] m Yoon Park [email protected] om TREK Sensititre Unscheduled Service As Required Technical support Preventive Maintenance Bi-Annual Tel: 1-800-267-6391 micro.catechsupport@thermofishe r.com VIDAS Check instrument temperature Daily, Weekly, Technologist Monthly Unscheduled Service As Required Biomerieux UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY

NOTE: This document is Uncontrolled When Printed. Any documents appearing in paper form that are not stamped in red "MASTER COPY" are not controlled and should be checked against the document (titled as above) on the server prior to use. Management System\UHN_Mount Sinai Hospital Microbiology\Quality Manual Policies and Procedures\Equipment\

Policy # QEQMI03001 Page 9 of 12 Department of Microbiology Quality Manual Version: 1.5 CURRENT Section: Equipment Subject Title: Equipment Maintenance Procedure

Equipment Maintenance Item Frequency Maintenance by Preventive Maintenance Scheduled Support Service Toll: 1-800-361-7321 [email protected] Virtuo Unscheduled Service As Required Biomerieux Preventive Maintenance Bi-Annual Biomerieux Support Service Toll: 1-800-361-7321 [email protected] Vitek  Check Reader Temperatures Daily Technologist  Check Reader Status  Check Backup  Swap tape  Check and clean arrays Weekly QC Technologist  Finalize cultures older than 7 days  Wash trays  QC susceptibility cards upon arrival of new lot# and weekly  QC identification cards upon arrival of new lot# Preventive Maintenance Bi-Annual Biomerieux Support Service Toll: 1-800-361-7321 [email protected] Vitek MS Monitoring / Fine tuning Fine Weekly Technologist & Tuning Biomerieux Desiccant check Adapter cleaning Monthly Technologist O ring / seal check

Replace Detector Annual Biomerieux /As Required Support Service Unscheduled Service As Required Toll: 1-800-361-7321 Preventive Maintenance Annual [email protected] WASP Routine maintenance Daily, Monthly Specimen management Unscheduled As Required COPAN Diagnostics

UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY

NOTE: This document is Uncontrolled When Printed. Any documents appearing in paper form that are not stamped in red "MASTER COPY" are not controlled and should be checked against the document (titled as above) on the server prior to use. Management System\UHN_Mount Sinai Hospital Microbiology\Quality Manual Policies and Procedures\Equipment\

Policy # QEQMI03001 Page 10 of Department of Microbiology 12 Quality Manual Version: 1.5 CURRENT Section: Equipment Subject Title: Equipment Maintenance Procedure

Equipment Maintenance Item Frequency Maintenance by Preventive Maintenance Bi-Annual Tel: 905-460-7053 Tech Support Toll: 1-877-927-7457 [email protected] WASPLAB Routine maintenance Daily, Weekly, Technologist Monthly Unscheduled As Required COPAN Diagnostics Preventive maintenance Bi-Annual Tel: 905-460-7053 Tech Support Toll: 1-877-927-7457 [email protected] Water deionizer Check Resistivity Daily Technician Check bacterial count Weekly QC technologist

Record all maintenance work done on Routine Maintenance Form in Analyzer History Logs per equipment.

Related Documents:

Equipment Log QEQMI02001 Analyzer History Log Table of Contents QEQMI3001h Equipment Maintenance Procedure QEQMI03001

UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY

NOTE: This document is Uncontrolled When Printed. Any documents appearing in paper form that are not stamped in red "MASTER COPY" are not controlled and should be checked against the document (titled as above) on the server prior to use. Management System\UHN_Mount Sinai Hospital Microbiology\Quality Manual Policies and Procedures\Equipment\

Policy # QEQMI03001 Page 11 of Department of Microbiology 12 Quality Manual Version: 1.5 CURRENT Section: Equipment Subject Title: Equipment Maintenance Procedure

Record of Edited Revisions

Manual Section Name: Equipment Maintenance Procedure

Page Number / Item Date of Revision Signature of Approval Annual Review March 1, 2004 Dr. T. Mazzulli Annual Review March 1, 2005 Dr. T. Mazzulli Annual Review July 23, 2006 Dr. T. Mazzulli Annual Review April 26, 2007 Dr. T. Mazzulli Annual Review June 16, 2008 Dr. T. Mazzulli Annual Review June 16, 2009 Dr. T. Mazzulli Annual Review June 11, 2010 Dr. T. Mazzulli Annual Review May 31, 2011 Dr. T. Mazzulli Annual Review May 31, 2012 Dr. T. Mazzulli Updated list May 31, 2012 Dr. T. Mazzulli Annual Review May 31, 2013 Dr. T. Mazzulli Update list July 30, 2014 Dr. T. Mazzulli Include:Vitek MS, Rotorgene, Simplexa, Epimotion, Evolis, EasyMag, GeneXpert, Chef II/III, Max Q4000 Shaker, GelDoc XR+ Remove: AXSYM, Magnapure, P-lab Annual Review July 30, 2014 Dr. T. Mazzulli Annual Review July 20, 2015 Dr. T. Mazzulli Updated with Biorad July 20, 2016 Dr. T. Mazzulli Annual Review Incubator O2 and CO2 updated to Biweekly cleaning January 12, 2017 Dr. T. Mazzulli schedule including:  Incubator  Shelves  Humidity pan Replace distilled water in humidity pan Added Virtuo maintenance Added Cobas 6800 Maintenance May 31, 2017 Dr. T. Mazzulli Remove BacT/Alert Maintenance , Rotorgene, Simplexa Annual Review Annual Review April 24, 2018 Dr. T. Mazzulli TREK Sensititer , WASPLAB and BioFire Added. Minor format change August 10, 2018 Dr. T. Mazzulli UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY

NOTE: This document is Uncontrolled When Printed. Any documents appearing in paper form that are not stamped in red "MASTER COPY" are not controlled and should be checked against the document (titled as above) on the server prior to use. Management System\UHN_Mount Sinai Hospital Microbiology\Quality Manual Policies and Procedures\Equipment\

Policy # QEQMI03001 Page 12 of Department of Microbiology 12 Quality Manual Version: 1.5 CURRENT Section: Equipment Subject Title: Equipment Maintenance Procedure

Page Number / Item Date of Revision Signature of Approval Addition of Altostar January 22, 2019 Dr. T. Mazzulli Annual Review April 26, 2019 Dr. T. Mazzulli Annual Review May 25, 2020 Dr. T. Mazzulli Full document review included in all updates. Bi-annual review conducted when no revision had been made within 2 years. Page Number / Item Date of Revision Edited by: Added note for unscheduled service “will label equipment Jan 12, 2021 Dorna Zareianjahromi as “Out of service” or “not in use” as appropriate” Minor formatting change April 08, 2021 Jessica Bourke Updated Seegene Starlet contacts April 24, 2021 Jessica Bourke Updated seegene May 28, 2021 Wayne Chiu

UNIVERSITY HEALTH NETWORK/MOUNT SINAI HOSPITAL, DEPARTMENT OF MICROBIOLOGY

NOTE: This document is Uncontrolled When Printed. Any documents appearing in paper form that are not stamped in red "MASTER COPY" are not controlled and should be checked against the document (titled as above) on the server prior to use. Management System\UHN_Mount Sinai Hospital Microbiology\Quality Manual Policies and Procedures\Equipment\